background image

DPM3-E Electrical User Manual, 1st Edition

User Manual - DPM3-E Series Electrical Panel Meters

26

Technical Specifications

Technical Specifications

Input

DC Voltage

AC Voltage

DC Current

AC Current

Range....Input 

Impedance

2V....100kΩ

20V....1MΩ

200V....1MΩ

600V....1MΩ

2V....75kΩ

20V....850kΩ

200V....850kΩ

600V....850kΩ

200mA....0.75 Ω

1A....0.014 Ω

5A....0.014 Ω

50mV.... 1.8 MΩ

60mV....1.8 MΩ

100mV....1.8 MΩ

200mA.... 0.75 Ω

1A....0.014 Ω

5A....0.014 Ω

50mV.... 1.5 MΩ

60mV....1.5 MΩ

100mV....1.5 MΩ

Input Frequency 

Range

40Hz to 10kHz True RMS 

Measurement

40Hz to 10kHz True RMS 

Measurement

Resolution

2V....0.1 mV

±20V....1mV

±200V....10mV

±600V....0.1 V

2V....0.1 mV

20V....1mV

200V....10mV

600V....0.1 mV

200mA....0.01 mA

1A....1mA

5A....1mA

50mV....0.01 mV

60mV....0.01 mV

100mV....0.01 mV

200mA....0.01 mA

1A....1mA

5A....1mA

50mV....0.01 mV

60mV....0.01 mV

100mV....0.01 mV

Accuracy

2V.... 0.05% rdg  

± 0.3 mV

±20V....0.05% rdg 

± 3mV

±200V....0.05% rdg 

± 30mV

±600V....0.05% rdg 

± 0.3 V

2V.... 0.3% rdg ± 0.3 mV

20V....0.3% rdg ± 3mV

200V....0.3% rdg ± 30mV

600V....0.3% rdg ± 0.3 V

200mA.... 0.1 % rdg ± 0.05 mA

1A....0.1 % rdg ± 5mA

5A....0.1 % rdg ± 5mA

50mV.... 0.1 % rdg ± 0.1 mV

60mV....0.1 % rdg ± 0.1 mV

100mV....0.1 % rdg ± 0.1 mV

200mA.... 0.3 % rdg ± 0.05 mA

1A....0.3 % rdg ± 5mA

5A....0.3 % rdg ± 5mA

50mV.... 0.3 % rdg ± 0.1 mV

60mV....0.3 % rdg ± 0.1 mV

100mV....0.3 % rdg ± 0.1 mV

Accuracy 

Conditions

AC or DC Input:

15 minutes warmup

25°C  ± 5°C ambient temperature

100 ppm/°C temperature coefficient

10-75% RH non-condensing 

AC Input:

3% to 100% of input range

45 Hz to 400 Hz sine wave

40 Hz to 10 kHz Ac/-(1% + 20 digits)

Crest Factor: 3, Ac/-(0.2% + 10 digits); 5, Ac/-(1% 

+ 20 digits)

Conversion

Technique

Sigma-Delta

Resolution

±15 bits

Conversion rate

20 times per second

Display

Range

-19999 / +19999, 5 LED digits 14mm (Programmable color Red, Green, Amber)

LEDs

8, functions and outputs status

Display refresh rate

20 times per second

Display / Input 

overrange indication 

"-

oUEr

” , “

oUEr

Relays 

-A2R Only

(2) Relays, Form C 

SPDT

Nominal contact rating....................8A at 250VAC / 24VDC

Maximum switching current (resistive load)....................8A

Maximum switching power.........................2000VA / 192W

Maximum switching voltage...................400VAC / 125VDC

Contact resistance...........................≤100mΩ at 6VDC at 1A

Operate time.................................................................≤10ms

Analog Output  

-A2R Only

Type

4-20 mA Sourcing

Maximum load

≤500Ω

Resolution

13 bits

Accuracy 

0.1%FS ±1 bit

Response time

10ms

Thermal drift

0.5µA / ºC

Содержание ProSense DPM3-E Series

Страница 1: ...ProSense DPM3 E Series Digital Panel Meter for true rms ac or dc voltage and amperage electrical input signals Models User Manual DPM3 E H DPM3 E L DPM3 E A2R H DPM3 E A2R L TRMS TRMS...

Страница 2: ...7 Installation 7 Wiring Terminals 8 Wiring Examples 10 DC AC Voltage Input Wiring Ranges 2 20 200 600 V 10 DC AC Current Input Wiring Ranges 1A 5A 10 DC AC Current Input Wiring Range 200mA 10 DC AC C...

Страница 3: ...17 Logic functions 17 Logic functions menu 18 Table of programmable functions 19 Configuration Lock Out 20 Lock out menu diagram 21 Lock out menu diagram continued 22 Output Options 23 Relay Configur...

Страница 4: ...pean legislation regarding waste of electrical and electronic equipment from products put on the market after August 13 2005 the user can give it back without any cost to the place where it was acquir...

Страница 5: ...manuals the best in the industry We rely on your feedback to let us know if we are reaching our goal If you cannot find the solution to your particular application or if for any reason you need techn...

Страница 6: ...n parameters can be totally or selectively locked out to prevent unauthorized or accidental changes to the meter s operation Other features include memory and reset of minimum valley and maximum peak...

Страница 7: ...g the unit in place put the fixing clips on both sides of the case and slide them through the guide tracks until they reach the panel at the rear side Press slightly to fasten the clips to the latchin...

Страница 8: ...2 3 4 1 2 3 4 5 6 CN4 CN5 CN6 1 2 CN2 Electrical Inputs 1 Common 2 Shunt 2V 3 200mA 4 1A 5A 5 20 200 600V CN3 1 Common 2 Peak 3 Valley 4 Hold Polarity insensitive for DC power CN1 AC Supply DC Supply...

Страница 9: ...ts CAUTION The isolation rating of the input common of the meter with respect to the option card common and to the Logical Functions common if used is 250 Vrms and 500 Vrms with respect to AC Power te...

Страница 10: ...es 1A 5A Terminals CN2 1 2 3 4 5 I N L 1 5 A DC AC DC AC Direct Connection Connection using and AC Current Transformer Terminals CN2 1 2 3 4 5 I N L 1A 5A AC MAX R1 R2 Important When using a current t...

Страница 11: ...reading AC using true RMS 2 MAX Indicates peak displayed 3 MIN Indicates valley displayed 4 PROG Indicates programming mode 5 DISPLAY Displays the input variable Displays programming parameters 6 UP D...

Страница 12: ...s visualization mode the meter automatically switches back to RUN mode after 15 seconds sine the last key press 1 Input Menu 2 3 4 5 Display Menu Relay Configuration Menu Analog Output Menu Logic Func...

Страница 13: ...configuration data is stored in memory The meter will then return to RUN mode Input Configuration The figure below shows the input configuration menu Divided into four submenus each one of them separ...

Страница 14: ...ximum in both ScAL and tEAcH methods ENTER Vertical displacement UP Changes active digit SHIFT Horizontal displacement Programming Key Pad Scaling For linear signals scaling is achieved by programming...

Страница 15: ...ts of the process is known tEACH method The input values are introduced directly from actual electrical values The input signal device must be connected to the meter and operational when each point is...

Страница 16: ...ivates the function Measured at 230V AC power supply display 100 00 amber colour and with no options Additional Functions Several functions can be controlled via Key Pad that will produce different ac...

Страница 17: ...en stored in the non volatile memory of the meter Factory configuration INPUT 600V DC DISPLAY Input 1 0000 0 Display 1 0000 0 Input 2 600 0 Display 2 600 0 Filter P 0 Round 01 Brightness High DISPLAY...

Страница 18: ...meter internal operation nor the analog and relay outputs if present Logic functions diagram 1 2 3 4 1 2 3 4 CN3 1 2 3 4 PIN INPUT Factor Default Functions Function PIN 1 COMMON PIN 2 INP 1 PEAK Funct...

Страница 19: ...he peak value MAX Low level 4 VALLEY Displays the valley value MIN Low level 5 RESET PEAK VALLEY Perform a reset of the peak or the valley depending on which is being displayed Falling edge 6 HOLD Fre...

Страница 20: ...ecommend changing the code set at factory and to write down your personalised code and keep it in a safe place TOTAL LOCKOUT The access to the programming routines to read data is allowed even if all...

Страница 21: ...er will return automatically to the run mode Total lockout programming is achieved by changing the totLC variable to 1 changing it to 0 will lead to the selective lockout of the programming variables...

Страница 22: ...M3 E Series Electrical Panel Meters 22 Lock out menu diagram continued 0 allows its programming 1 locks the access to programming Only appear if the corresponding options are present in the meter InPu...

Страница 23: ...wing relay operational parameters must also be set a HI LO ACTING MODE In HI mode the output activates when the display value exceeds the setpoint level and in LO mode the output activates when the di...

Страница 24: ...git SHIFT Horizontal displacement Programming Key Pad Direct access to the relay setpoints value programming It is possible to directly access the relay setpoint values without the need to go through...

Страница 25: ...riety of terminal equipment such as graphic recorders controllers remote displays or other devices that accept input data in analog form The display values producing the full scale output OUT HI and O...

Страница 26: ...0 1 mV 60mV 0 1 rdg 0 1 mV 100mV 0 1 rdg 0 1 mV 200mA 0 3 rdg 0 05 mA 1A 0 3 rdg 5mA 5A 0 3 rdg 5mA 50mV 0 3 rdg 0 1 mV 60mV 0 3 rdg 0 1 mV 100mV 0 3 rdg 0 1 mV Accuracy Conditions AC or DC Input 15...

Страница 27: ...sing 0 5A 250V DIN 41661 22 53 VAC 50 60 Hz 10 5 70 VDC recommended fusing 2A 250V DIN 41661 Power Consumption 5W without options 8W max Filter Cutoff frequency 4Hz to 0 05Hz Slope 20dB Dec Environmen...

Отзывы: