You may want to adjust an offset value also, see separate OFFSET page for this feature.
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Scale Factor adjustment
After you have calibrated your meter, you can use the SCALE feature to make fine adjust-
ments to calibration, without affecting the calibration itself.
Example
Changing volume units of measure from litres to Imperial gallons
You could also use the SCALE to convert your readout from litres to imperial gallons, with-
out affecting the calibration. Simply set SCALE = 0.220 and your meter which was cali-
brated in litres will now read in imperial gallons.
Changing weight units of measure from kg to pounds
You could also use the SCALE to convert your readout from kg to pounds, without affecting
the calibration. Simply set SCALE = 2.205 and your meter which was calibrated in kg will
now read in pounds.
Lockout Switch must be OFF
Press 3 seconds
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Press repeatedly until you see
SCAL
, followed by the existing scale
factor.
(Default =
1.000
)
1
2
3
4
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Use DIGIT button to select each digit
in turn, UP or DOWN buttons to
increase or decrease each digit’s
value. Press OK when done.
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Done!
!! This feature is available in Advanced Mode only !!