21
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Done!
5
6
7
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
The display will show
SE. 01
Apply your 1st known load to the system
and press OK.
The display will now show
rd 01
You can now tell the meter what it should
display for this load.
Select each digit in turn with the DIGIT
button, and increase or decrease each
digit’s value using the UP or DOWN
button, until the display is set as required.
Press OK when done.
The display will show
Ad.Pt
Press OK if you want to Add another
calibration point, you will repeat the previ-
ous steps, each time incrementing “Set
XX” and “rd XX” up to Set 09 and rd 09
which are the last available points.
If you want to finish adding points, press
Set2 until the display shows
SAVE
, and
then confirm by pressing OK
If you want to abort the setup, press Set2
until the display shows
quit
and then
press OK
Linearisation with Direct Calibration - cont’d.
Input
Display
Set00
Set01
Set02
Set03
Set04
Set05 Set06
Set07
Set08
rd 08
rd 07
rd 06
rd 05
rd 04
rd 03
rd 02
rd 01
rd 00