Guidance and Manufacturer's Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The
customer and/or user of this device should ensure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test
level
Compliance level
Electromagnetic
environment -guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
relative humidity
should be at least
30%.
Electrical Fast
Transient /
Burst
IEC 61000-4-
4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV Line to Line
±1 kV Line to Line
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
0% UT in 0.5 cycle
at 0°, 45°, 90°,
135°, 180°, 225°,
270°, and 315°
0% UT in 1 cycle at
0°
70% UT in 25/30
cycles at 0°
0% UT in 250/300
cycles at 0° and
180°
0% UT in 0.5 cycle
at 0°, 45°, 90°,
135°, 180°, 225°,
270°, and 315°
0% UT in 1 cycle at
0°
70% UT in 25/30
cycles at 0°
0% UT in 250/300
cycles at 0° and
180°
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power
Frequency
(50/60 Hz)
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels