14
Alarms
The Infant Oximeter Box is equipped with an LED indicator that will blink red when a
high priority alarm occurs and blink yellow for medium priority alarm.
For more information about the alarm system, refer to the Instructions for Use of
Guardian Angel® Rx Digital Vital Sign Monitoring System.
Care and Maintenance
The advanced digital circuitry within the device requires no calibration or periodic
maintenance. Field service or repair of this system is not possible. Do not attempt to
open the case of the device for that will cause damage and void the warranty. If the
device is not functioning properly, see “Troubleshooting” section for more
information.
Cleaning and Disinfection
The device is expected to be cleaned and disinfected before use, i.e., once per day. It
has been tested to withstand 300 times of cleaning within 1.5 years of re-use life. Do
not use the device that has exceeded its re-use life.
We recommend you clean and disinfect the device with the instructions below.
First, lightly wipe the surface of the device with a soft cloth dampened with rubbing
alcohol for cleaning. Secondly, disinfect the surface of the device with a soft cloth
saturated with a solution of 10% chlorine bleach in tap water. Lastly, allow the device
to dry thoroughly. Visual inspection is necessary at the end of cleaning and
disinfecting cycle. Repeat the previous cycles to remove visible residual soil on the
device. Do not use a visibly soiled device again.
Reuse Life
This device is reusable with an expected life of 1.5 years. However, if you notice any
signs of deterioration from below, stop using it and replace it with a new one or
contact Aulisa Customer Support by going online at www.aulisa.com:
-
button malfunctions;
-
cracks appear on external case;
-
edge of the gel covering the sensor probe window starts curling up;