12
Device Pairing
Automatic Pairing
The system automatically scans and pairs to the Aulisa sensor module(s) from the
same starter kit.
Manual Pairing
Follow the below instructions to manually setup pairing of a new Infant Oximeter
Module.
Step 1:
Turn on the Infant Oximeter.
Step 2:
In the Setting menu, select “PAIRING”
→
“SENSOR MODULE”.
Step 3:
Scan the QR Code or key in the serial number located on the back of the
Infant Oximeter Box.
Step 4:
Press “CONFIRM” if the serial number (SN) displayed matches with the one
on the Infant Oximeter Box.
Step 5:
To confirm that the process was successful, ensure that the Bluetooth
connection status icon on the MAIN screen is lit blue.
NOTE:
The device must be placed within 32.8 feet (10 meters) to the Guardian Angel®
Rx Digital Vital Sign Monitoring System.
NOTE:
The Bluetooth connection status icon will turn blue once the pairing succeeds.
NOTE:
The power LED on the Infant Oximeter Box will blink green when pairing
succeeds, and data transmission starts.
NOTE:
Up to four (4) Infant Oximeter can be stored in the system.