MetaFridge

TM

 CFD-50

 

User’s Manual 

 

 

All material that is labeled with 

⚠ 

may affect either product safety or user’s safety and should be handled with 

caution.         

 

 

⚠   

Please read the manual carefully before use 

Keep the manual in a safe place for future reference

 

For safety purposes, please remove all package materials before use. 

Содержание MetaFridge CFD-50

Страница 1: ...is labeled with may affect either product safety or user s safety and should be handled with caution Please read the manual carefully before use Keep the manual in a safe place for future reference Fo...

Страница 2: ...1 5 1 Temperature Display and Status LED 11 5 2 30 DTR Daily Temperature Record Display 12 5 3 Alarms 16 5 4 USB Download and 30 DTR reports 16 6 Maintenance 17 6 1 Daily maintenance tasks 17 6 2 Annu...

Страница 3: ...ompartment Integrated power management system defending against voltage spikes and brown outs Integrated voltage stabilization to utilize a wide range of input voltages 110 230 VAC Normal operation in...

Страница 4: ...hamber door gaskets D 30 DTR display E Temperature display F Water tank lid latch G Water tank lid H Electronics enclosure I USB port J Condenser guards K Vaccine chamber door latch L Lifting handles...

Страница 5: ...5 P External condenser Q Thermal safety shut off mechanical thermostat R Water tank lid hinge...

Страница 6: ...ation to prevent damage c Keep door closed when moving the CFD 50 d If the CFD 50 needs to be moved after initial installation remove the water from the water tank using a siphon hose before moving e...

Страница 7: ...d length is 2 meters d Position the CFD 50 away from direct sunlight and any source of heat e g gas stove fireplace etc e CFD 50 must not be placed in or around flammable or corrosive gases that could...

Страница 8: ...50 Install the funnel securely in the fill hole Fill a bucket with clean drinking water If there is no local source of clean drinking water use bottled water Step 3 Pour 32 L of water into the water...

Страница 9: ...ove excess water using a piece of tubing as a siphon Adjust the water level by adding removing water until the water level is at the bottom of the fill tube Too little water will reduce holdover time...

Страница 10: ...tep 4 Do not use the CFD 50 until the display indicates that the vaccine storage temperature is below 8 C and the 30 DTR indicates at least 2 days of remaining holdover which usually takes 6 hours wit...

Страница 11: ...peration to ensure the safe use of the CFD 50 Only approved vaccines or medical supplies should be stored in the refrigerator Do not store any foods beverages or flammable explosive volatile items suc...

Страница 12: ...dover current power availability and ongoing alarms 30 DTR display and navigation buttons Navigation buttons Menu button images Tap the Menu button to move between CURRENT HISTORY and ALARM screens Ba...

Страница 13: ...d over time and needs to be replaced Plug icon button The power icon indicates whether mains power in the CFD 50 s usable range 82 290 V is currently available Alarm history indicators Dots indicate v...

Страница 14: ...rent active screen Date The date currently being viewed An arrow towards the bottom of the screen indicates which of the past 30 days is being viewed Temperature history For each day the minimum avera...

Страница 15: ...emperature excursion threshold is 10 hours above 8 C the alarm will begin 10 hours after the temperature first exceeds 8 C Separate alarms are created when an alarm spans multiple days an alarm that b...

Страница 16: ...ondition has ended The Active Alarm screen shows the alarm type and associated information Pressing the Menu button will show the CURRENT and HISTORY screens and allow normal navigation After the disp...

Страница 17: ...urfaces using a soft cloth dipped with neutral non corrosive cleaning agent Do not use any brushes acidic or alkali detergents Use a dry cloth to wipe clean surfaces Do not use any organic solvents ho...

Страница 18: ...and any inflammable gases h Do not touch any electrical components such as plugs switches etc with wet hands in order to avoid electric shock i Do not damage or break the power cord and power plug or...

Страница 19: ...oltage input requirements Refrigerator cabinet temperature too high Prolonged door opening or door ajar Avoid prolonged door openings and make sure door is closed after use Chamber may take some time...

Страница 20: ...longed door openings and make sure door is closed after use Refrigerator is not positioned on a leveled and stable floor Position the refrigerator on a stable floor and level it so that condensate flo...

Страница 21: ...Ambient Operating Temperature and Humidity Range 10 43 Relative Humidity 95 Vaccine Storage Chamber Temperature Range 2 8 Holdover Time 43 C 120 hr 5 day Vaccine Storage Capacity 50 L Input Voltage R...

Страница 22: ...we will provide the following coverage to you 1 Two year warranty on the device included at no extra cost 2 Under warranty coverage the manufacturer will repair or replace device components that have...

Страница 23: ...ucma Global Medical Co Ltd ADD NO 315 Qianwangang Road Economic Technological Development Qingdao China Zip 266000 Tel 86 0532 88965799 Fax 86 0532 88965766 Email info aucma com cn Website www aucmagl...