Page 46
Press the ‘Autoapproach Details’ where the following parameters should be
entered:
Coarse Axis:
Axis 3
Coarse Dir.:
Forward
Coarse Device:
ANC150
Coarse Trig. Pol.:
High Active
Coarse Trig. Hold Time:
1ms
Check the ‘Axis 3’ settings to have reasonable values for the coarse
positioning, for example:
Frequency:
200-1000Hz
Amplitude:
30V (@RT), 50V(@LT)
3. Z Spectroscopy:
After the ‘Auto Approach’, run a so-called ‘z Spectroscopy’. This is a
spectroscopy where the feedback parameter is monitored while the z
scanner moves the sample surface towards the tip until some breakdown
condition is reached. To perform this spectroscopy, we recommend to enter
the following values/ranges into one of the three ‘Spec’ windows
(Spec1,Spec2 or Spec3):