Thanks for using our products, please read this manual thoroughly before 
operation. 

MODEL:

 

 regulated DC power supply is a high reliability, 

including both stable voltage and stable current function, 

TPR3003T/TPR3005T

Restricted current lower voltage, short 

circuit protection, thermal protection with load ability etc

The power supply is suitable for the 

laboratory, electronics, communications equipment maintenance, Product line, scientific research 
and teaching units. If power working long hours running state, overheating will shut off the output, 
when machine temperature returned to normal, output power restore default state. The machine 

OVERVIEW

SPECIFICATION

1. Rated working conditions and external dimension

Input AC voltage

    

AC 220V

5%, Frequency 50Hz

    

(If according to the requirement can be designed in 110V

5% 60Hz)

The working conditions

    Temperature rang: 0

~40

    Relative humidity less than 80%

Storage conditions

    Temperature rang: -20

~80

    Relative humidity less than 80%

Shape dimension

    (L)270

(W)125

(H)170mm

Accessories

    users manual    1pc 
    power cable

      

1pc

2. Technical specification

Output voltage from 0 to 30V between nominal continuous adjustable.
Output current from 0 to nominal value between continuous adjustable.

                                CV less than 0.01

+1mV

                                CV less than 0.2

+1mV

    

Power regulation

INSTRUCTIONS

1.

 

Note

(1) AC input

AC input should be 220V

10%, 50Hz

 

(If you are in the box 110V 10% 60Hz behind can indicate)

(2) Scattered heat

Do not exceed 40 degrees Celsius temperature environment in the local use, fan in the posterior 
instrumentation, should have enough space, so scattered heat.

(3) Output voltage overshoot

When switching power supply device, the output voltage is not more than preset value.

(4) Overheating protection

This series of power is the perfect overheating protection function, can work under state 
protection power will not lead to internal components overheating and burned power and user 
equipment. If power transformer, internal temperature overheating will automatically cut off the 
input voltage, When the internal temperature of power to the normal range, power will restore
default working state.

2. limiting Current Settings

(1)The decision to power instrument maximum safety.
(2)

 Using a wire temporarily voltage output terminals (+) and (-) terminal circuit connection.

(3)
(4)

Rotating voltage control knob, output voltage regulating 3V to 5V

    

Rotating voltage control knob, until the CC designator light.
Adjust current knob to the current needs

Load regulation:

  CV less than 0.01

+5mV     

I

3A

load

                                    CC less than 0.2

+5mA       

I

3A

load

                                    CV less than 0.01

+7mV      I

3A

load

                                    CC less than 0.2

+7mA       

I

3A

load

Ripple and noise:  

CV

0.5mV

I

3A

RMS                                              

load

                                    CC

3mA

I

3A

RMS                                                    

load

    

 (5Hz-1MHz)  

       CV

1mV

I

3A

RMS                                                    

load

                                    CC

5mA

I

3A

RMS                                                    

load

 

Voltage accuracy: 

 0.5%rdg+2byte

 

Current accuracy: 

0.5%rdg+2byte

 

Display resolution:

0.5%rdg+2byte

 

  

High voltage test:

 

Leakage current  I

1 mA 

(Test conditions: 1700VAC/2s From the input 

    power to the ground, Input high AC voltage: 1700V, Time: 2 seconds).

Insulation resistance: 

Insulation resistance more than 100M

 

(Test conditions: 500V /5s 

dc

    

From the input power to the ground, Input high DC voltage: 500Vdc, Time: 5 seconds).

has perfect protection function, can protect user equipment are not consumed.