29
Signal Generator
8.
Press the
Ampl Stop
soft function key, and input 0dBm.
It will change the stop amplitude electric level of the step sweep.
9.
Press the
Points
soft function key, and input 9.
It will set the sweep points as 9.
10. Press the
Step Dwell
soft function key, and input 500mse.
11. It will set the dwell time of each point as 500 ms.
12. Press the
Return
to return to the last operation menu, press the
Sweep
soft function key, and press the
Freq & Ampl
soft function key.
It will set the step sweep as frequency data sweep as well as amplitude data
sweep. This soft function key can be selected to return to the last menu and run
the sweep function.
13. Press
RF On/Off
.
The status of the indicator on the display screen will be changed from RF OFF
into RF ON.
14. Press
Single Sweep
.
It will execute single sweep of frequency and amplitude configured in step
sweep, and the result is output to the RF OUTPUT connector.
2-2-2-3 Activate continuous step sweep
Press
Sweep Repeat Single Cont
.
It will switch sweep from single sweep to continuous sweep. At the moment,
continuous repeated frequency and amplitude configured in step sweep can be
obtained from the RF OUTPUT connector.
List sweep is used for building a list of ant frequency, amplitude and dwell time
values, and sweeping RF output based on the entries in the List Mode Values table.
Содержание Gratten GA1483
Страница 1: ......
Страница 57: ......
Страница 60: ...48 5 2 AMPT...
Страница 61: ...49 49 Signal Generator 5 3 FM...
Страница 62: ...50 5 4 FREQ 5 5 LF OUT...
Страница 63: ...51 51 Signal Generator 5 6 PULSE 5 7 RECALL...
Страница 64: ...52 5 8 SAVE...
Страница 65: ...53 53 Signal Generator 5 9 SWEEP...
Страница 66: ...54 5 10 SYSTEM...
Страница 74: ......