23
Signal Generator
2-2 Configure RF output
This section will introduce how to build RF output of continuous wave and
frequency sweep to you.
2-2-1 Configure RF output of continuous wave
Use these procedures to learn how to set the following parameters:
1.
RF output frequency
2.
RF reference and frequency offset
3.
RF output amplitude
4.
Amplitude reference and amplitude offset
2-2-1-1 Set RF output frequency
1. Press
PRESET
.
It makes the signal generator return to the status defined when the signal
generator leaves factory.
2. Observe the FREQUENCY region (it is in the left upper corner of the display
screen) in the display screen. The value displayed is the specified highest
frequency of the signal generator.
3. Press
RF On/Off
.
RF signals can be acquired in the RF OUTPUT connector by pressing the
RF On/Off
hard function key only. The status of the indicator on the display
screen will be changed from RF OFF(RF OFF) into RF ON(RFON), and at the
moment, the highest specified frequency will be output to the RF OUTPUT
connector (it is at the lowest power electric level of the signal generator).
4. Press the
FREQ
hard function key, and input 700 MHz.
At the moment, 700 MHz of RF frequency will appear in the FREQUENCY
Содержание Gratten GA1483
Страница 1: ......
Страница 57: ......
Страница 60: ...48 5 2 AMPT...
Страница 61: ...49 49 Signal Generator 5 3 FM...
Страница 62: ...50 5 4 FREQ 5 5 LF OUT...
Страница 63: ...51 51 Signal Generator 5 6 PULSE 5 7 RECALL...
Страница 64: ...52 5 8 SAVE...
Страница 65: ...53 53 Signal Generator 5 9 SWEEP...
Страница 66: ...54 5 10 SYSTEM...
Страница 74: ......