and downward sweep direction. If the
Sweep Direction Down Up
soft function
key is set as Up (UP), the values will be swept from the start frequency/amplitude to
the stop frequency/amplitude. When the Sweep Direction Down Up soft function
key is set as Down (Down), the values will be swept from the stop
frequency/amplitude to the start frequency/amplitude.
2-2-2-2 Configure and activate single step sweep
During the process, a step sweep with nine sweep points and the following
parameters will be built:
Frequency range is from 500MHz to 600MHz.
Amplitude range is from -20dBm to 0dBm.
Dwell time of each point is 500ms.
Detailed operation steps are as follows:
1.
Press
PRESET
.
2.
Press
SWEEP
.
It will open a sweep soft function key menu.
3.
Press
Sweep Repeat Single Cont
.
It will switch the sweep repeat mode from Continuous to Single.
4.
Press
Configure Step Sweep
.
5.
Press the
Freq Start
soft function key, and input 500MHz.
It will change the start frequency of the step sweep into 500MHz.
6.
Press the
Freq Stop
soft function key, and input 600MHz.
It will change the stop frequency of the step sweep into 600MHz.
7.
Press the
Ampl Start
soft function key, and input −20dBm.
It will change the start amplitude electric level of the step sweep.
Содержание Gratten GA1483
Страница 1: ......
Страница 57: ......
Страница 60: ...48 5 2 AMPT...
Страница 61: ...49 49 Signal Generator 5 3 FM...
Страница 62: ...50 5 4 FREQ 5 5 LF OUT...
Страница 63: ...51 51 Signal Generator 5 6 PULSE 5 7 RECALL...
Страница 64: ...52 5 8 SAVE...
Страница 65: ...53 53 Signal Generator 5 9 SWEEP...
Страница 66: ...54 5 10 SYSTEM...
Страница 74: ......