switches and control knobs. 

 

Adjust intensive and focus respectively, make the traces clear. 

 

Adjust  CH1  position  knob  and  trace  turning  potentiometer  to  parallel  the  trace  and  horizontal  scale  (use  screw 
driver to adjust trace turning potentiometer 4). 

 

Input the calibrated signal to CH1 input port by using 10:1 probe. 

 

Set the switch AC-GND-DC in AC state. A waveform as fig.3 will appear on screen. 

 

Adjust focus to achieve clearest figure. 

 

Regard to other signals, by adjust vertical attenuation switch and sweep time to desired position, and achieve clear 
figure. 

 

Adjust vertical and horizontal position knob, in order to read out the amplitude and time of waveform easily. 

* The above are the foundational operations of oscilloscope, the operation of CH2 and CH1 are the same. 
 

4.

 

Operation of dual channel 

Adjust Vertical mode to DUAL, and trace of CH2 will appear on screen (the same with CH1). At this time, a square 
wave displays on CH1 (come from output waveform of calibration signal), and only one straight line displays on 
CH2, since no signal through this channel. Now put the calibration signal to CH2 input, which is the same with CH1, 
set  AC-GND-DC  switch  to  AC,  adjust  vertical  position  11)  and  19)  to  make  both  channels  as  fig.4.  Release 
ALT/CHOP switch, (set in ALT). The signals of CH1 and CH2 display on screen alternatively, this setting is used for 
scanning the two signals of shorter sweep time. Press ALT/CHOP switch, (set in CHOP). Signals of CH1 and CH2 
display on  screen  separately  with  the  rate of 250 kHz,  this  setting  is  used  for  scanning  the  two  signals of  longer 
sweep  time.  When  in  the  operation  of  dual  channel,  (DUAL  or  +/-),  must  select  the  signal  of  CH1  or  CH2  by 
triggering switch of signal generator, which act as trigger signal. If the signal of CH1 and CH2 synchronize, the both 
waveform  stably  display.  In  reverse,  only  the  signal  of  trigger  signal  generator  can  stably  display;  if  TRIG/ALT 
switch depressed, both the waveforms stably display.   

 
5.    +/- Operation 

      By adjusting the setting of “Vertical modes switch” to “add”, the algebraic sum of CH1 and CH2 will be displayed, 
or algebraic difference if the switch CH2 INV is depressed. For gain the accurate values of both, the attenuation setting 
of the both channels must be consistent. Vertical position can be adjusted by “

 

Position

”. In consideration of the 

linear change of vertical amplifier, setting the knob in central position is recommended.   
 
 

6.    Selection of trigger source   

Properly  selecting  trigger  source  is  crucial  to  effectively  use  the  oscilloscope.  User should  acquaint  the  selection 
function and operation sequence of trigger source. 

 

MODE switch 

       

AUTO

: In mode of Auto, sweep generator generates freely a sweep signal without trigger signal. It turns to trigger 

sweep  when  there  is  trigger  signal.  Usually  observe  a  waveform  at  first  time,  set  in  “AUTO”,  and  after  a  stable 
waveform is observed, adjust other setting. Having set other control parts, switch should be set to “NORM” trigger mode, 
as this mode more sensible. It should use “AUTO” mode when test DC signal or small signal.   
     

NORM

:  Normal  mode,  sweeper  often  keeps  in  stationary  state,  no  trace  display  on  screen,  When  trigger  signal 

through the valve level, which is set by “: Trigger level switch”, scan one time. Then the sweeper turn back to stationary 
state till to be triggered next time. When dual channel display “ALT” & “NORM” sweep, it doesn’t display unless CH1 
and CH2 have sufficient trigger level.   
     

TV-V: 

when observe a full field TV signal in TV field, set MODE switch to TV-V, synchronize the field signal of TV 

signal, sweep time often set in 2ms/div (one frame signal) or 5ms/div (interleaved signal in one field two frame).   
   

  TV-H:

 synchronize the TV line with line signal of TV signal, sweep time is often 10us/div, display several signal 

waveform,  adjust  sweep  time  to  required  linage  by  using  the  fine  tune  knob.  The  synchronized  signal  send  to 

Содержание AT7328 Series

Страница 1: ...d high light Voltage of accelerated pole up to 2kV it can be able to display clear traces even under high speed sweep The alternative triggering function can observe different signal waveform in two frequencies TV signal synchronous function CH1 output 50 output signal in the rear panel can drive frequency counter directly or other instrument Z axial input Function of bright adjustable can add fre...

Страница 2: ...ce Transformer Coaxial Directional Coupler Detector Power Divider 7 Advanced Milli volt Meter 8 Regulated DC Power Supply 1 AT TPR Series Low Ripple CV Digital Display Power Supply Industrial Products 2 AT E Series Ultra Ripple High Precision CV CC Power Supply Research Education Military Products 3 AT SPS Series Compensation Linear Power Supply 4 AT KPS Series Switching Adjustable Power Supply 5 ...

Страница 3: ...DD Addition of Channel 1 and Channel 2 Interruption Repetitive Frequency Approx 250KHz Input Coupling AC GND DC Max Input Voltage 300Vp p AC Frequency 1KHz When probe sets in 1 1 maximum effective readout value is 40Vpp 14Vrms Sine wave When probe sets in 10 1 maximum effective readout value is 400Vpp 140Vrms Sine wave CMRR 50 1 50KHz sine wave If CH1 and CH2 have the same sensibility 1000 1 50KHz...

Страница 4: ...C AC peak AC not exceed 1KHz Sweep time 0 2 Sec0 5Sec DIV Accuracy 3 Trimming 1 2 5 panel graduation Sweep expansion 10 times 10MAG Sweep time accuracy 5 20n sec 50n Linearity 3 10MAG 5 No calibration at 20ns 50ns Horizontal System Position by 10MAG 2DIV in center CRT RF MICROWAVE INSTRUMENT SERIES Specifications Sensibility Refer to Vertical axis Bandwidth DC 500KHz X Y Mode X Y Phase difference ...

Страница 5: ...g factory open the package immediately as this item is received to check whether there is any spoilage during transportation if yes contact supplier or sender immediately 2 Check power voltage AC 110V 220V is available check the voltage switch whether conforms with the local supply or not before switch on Notice Failure in select power will result in damage the oscilloscope Warning To prevent elec...

Страница 6: ...V peak Warning To prevent damage do not exceed these values Frequency of Max input voltage must be lower than 1 kHz If an AC volt overlaps on DC volt the maximum voltage of CH1 and CH2 input cannot exceed 300V therefore for a AC which average voltage is zero its peak peak value is 600Vpp Operation methods 1 Features of front panel See fig 1 CRT 6 Power Main power switch LED 5 lights when this swit...

Страница 7: ...l and trigger signal of CH2 invert simultaneously Triggering 24 External trigger input Used for trigger for external signal For this case Switch 23 it should be set in EXT 23 Selection of trigger source Trigger of internal INT or external EXT CH1 Choose CH1 as internal trigger signal source for vertical mode switch 14 is set in DUAL or ADD CH2 Choose CH2 as internal trigger signal source for verti...

Страница 8: ...al and horizontal deflection factor of oscilloscope 15 GND Grounding terminal of oscilloscope case 2 Features of rear panel See fig 2 34 AC Power AC Power input jack AC power wire connects here 35 Supporter For supporting the oscilloscope and lead out the power cord Operation methods 3 Basic operation Single channel operation Make sure whether the voltage of the unit is in accordance with your loc...

Страница 9: ...only the signal of trigger signal generator can stably display if TRIG ALT switch depressed both the waveforms stably display 5 Operation By adjusting the setting of Vertical modes switch to add the algebraic sum of CH1 and CH2 will be displayed or algebraic difference if the switch CH2 INV is depressed For gain the accurate values of both the attenuation setting of the both channels must be consi...

Страница 10: ...gger switch set in rising edger triggered and falling edge is triggered for See fig 6 6 Fig 6 Positive zone Level Trigger alternative switch When vertical mode set in dual channel display the switch used in alternative trigger and alternative display suitable for the mode of summarize with CH1 CH2 In alternative mode trigger signal alternates one time for every sweep period This mode is avail for ...

Страница 11: ...on signal adjust compensation capacitor to the best waveform no overshoot fillet upwarping See fig 9 9 Fig 9 Appropriate compensation Overcompensation Undercompensation 11 DC balance adjustment DC BAL Set the input coupling switch of CH1 and CH2 in GND trigger mode in AUTO adjust trace in centre position Adjust the attenuation switch between 5mV and 10mV tune DC BAL until the trace is below zero h...

Страница 12: ...r sampling Trigger switch Z Z axial input High voltage X X axial signal 1 CH1 Sweep generator Line scan switch 2 CH2 Trigger input amplify Trigger signal generator Line scan output amplify External Auto circuit Free oscillate signal To circuits of other units Power circuit 50 60Hz AC 50 60Hz Power trigger 2Vp p 1KHz 2Vp p 1KHz square waveform Calibration signal generator All specifications are sub...

Отзывы: