oscilloscope must be negative. (See fig 5) 

 

Function of trigger source: 

In order to display a stable waveform on screen, a signal which is relate with display signal in time should be provided 

to trigger circuit, the trigger switch is just used to select the trigger signal.   

CH1/CH2:

 Internal trigger mode in most cases. The signal which is sent to vertical input , separate one part to trigger 

circuit before pre-amplify. As the trigger signal is just the signal under tested, a stable waveform will display on screen. 

In DUAL or ADD mode, trigger signal is selected by trigger source switch.   
LINE

: Use frequency of AC supply as trigger signal. This method is very available to test the signal relate with power 

frequency. Such as AC noise of acoustic equipment, thyristor circuit etc.   

EXT

: Drive scan trigger circuit by external signal. Since the external signal has certain time relate with the signal under 

tested, the waveform can be display more independently. 
 
 
 
 
 
 
 

 

Trigger level and polarity switch 

It generates a sweep trigger signal when trigger signal through a preset valve level. The level can be changed by 
selecting trigger level, valve level move to positive when adjust toward “+”. And valve level move to negative when 
adjust toward “-”, and valve level set in average of signal when in centre. About positive signal, starting phase is 
variable.

  Note:  if  the  trigger  level  is  adjusted  over-positive  or  over-negative,  there  is  also  no  sweep  signal 

generates, because by now the trigger level has been exceed the amplitude of synchronized signal. 

When the 

polarity trigger switch set in “+”, rising edger triggered, and falling edge is triggered for “-”. (See fig 6) 

 

6    Fig 6               

Positive zone               

Level   

       

 

Trigger alternative switch 

When  vertical  mode  set  in  dual  channel  display,  the  switch  used  in  alternative  trigger  and  alternative  display 
(suitable for the mode of summarize with CH1, CH2).   

        In  alternative  mode,  trigger  signal  alternates one  time  for  every  sweep period. This  mode  is  avail  for  the test  of 
waveform amplitude and period, even the two waveform has no relative can be observed, but not suitable for measuring 
of phase versus time. About this measurement, both channels must be trigger by the source synchronize signal.   
      In dual  channel  display,  if  “CHOP”:&  “TRIG.ALT”  depressed  simultaneously,  cannot  synchronize  display,  due  to 
“CHOP” signal become trigger signal. Recommend to use “ALT” mode or directly select CH1 or CH2 as trigger signal 
source.   

7.

 

Sweep rate control 

Adjust sweep rate knob, select the waveform number that you want to observe. If too many waveform display on 
screen, adjust  sweep  time faster, if only one period waveform display on screen, can slow down the sweep time. 
When the sweep rate is too fast, only a part of period signal can be observed. But there may be only one straight line 
on screen for a square signal. 

8.    Sweep extension 

      Extreme high sweep rate is required when observe a part of a waveform. But if the part want to be observed is far 
from starting point of sweep, the observed waveform may be out of screen. Now we need the sweep extension switch. 
When the sweep extension switch is depressed, the display range can extend 10 times. Now the sweep extension switch 
is (the value of “Sweep rate switch”) multiple by 1/10. e.g., 1uSec/div can be extended to 100nSec/div. 

7       

10

  10 for extension     

    adjust position knob that you 

Содержание AT7328 Series

Страница 1: ...d high light Voltage of accelerated pole up to 2kV it can be able to display clear traces even under high speed sweep The alternative triggering function can observe different signal waveform in two frequencies TV signal synchronous function CH1 output 50 output signal in the rear panel can drive frequency counter directly or other instrument Z axial input Function of bright adjustable can add fre...

Страница 2: ...ce Transformer Coaxial Directional Coupler Detector Power Divider 7 Advanced Milli volt Meter 8 Regulated DC Power Supply 1 AT TPR Series Low Ripple CV Digital Display Power Supply Industrial Products 2 AT E Series Ultra Ripple High Precision CV CC Power Supply Research Education Military Products 3 AT SPS Series Compensation Linear Power Supply 4 AT KPS Series Switching Adjustable Power Supply 5 ...

Страница 3: ...DD Addition of Channel 1 and Channel 2 Interruption Repetitive Frequency Approx 250KHz Input Coupling AC GND DC Max Input Voltage 300Vp p AC Frequency 1KHz When probe sets in 1 1 maximum effective readout value is 40Vpp 14Vrms Sine wave When probe sets in 10 1 maximum effective readout value is 400Vpp 140Vrms Sine wave CMRR 50 1 50KHz sine wave If CH1 and CH2 have the same sensibility 1000 1 50KHz...

Страница 4: ...C AC peak AC not exceed 1KHz Sweep time 0 2 Sec0 5Sec DIV Accuracy 3 Trimming 1 2 5 panel graduation Sweep expansion 10 times 10MAG Sweep time accuracy 5 20n sec 50n Linearity 3 10MAG 5 No calibration at 20ns 50ns Horizontal System Position by 10MAG 2DIV in center CRT RF MICROWAVE INSTRUMENT SERIES Specifications Sensibility Refer to Vertical axis Bandwidth DC 500KHz X Y Mode X Y Phase difference ...

Страница 5: ...g factory open the package immediately as this item is received to check whether there is any spoilage during transportation if yes contact supplier or sender immediately 2 Check power voltage AC 110V 220V is available check the voltage switch whether conforms with the local supply or not before switch on Notice Failure in select power will result in damage the oscilloscope Warning To prevent elec...

Страница 6: ...V peak Warning To prevent damage do not exceed these values Frequency of Max input voltage must be lower than 1 kHz If an AC volt overlaps on DC volt the maximum voltage of CH1 and CH2 input cannot exceed 300V therefore for a AC which average voltage is zero its peak peak value is 600Vpp Operation methods 1 Features of front panel See fig 1 CRT 6 Power Main power switch LED 5 lights when this swit...

Страница 7: ...l and trigger signal of CH2 invert simultaneously Triggering 24 External trigger input Used for trigger for external signal For this case Switch 23 it should be set in EXT 23 Selection of trigger source Trigger of internal INT or external EXT CH1 Choose CH1 as internal trigger signal source for vertical mode switch 14 is set in DUAL or ADD CH2 Choose CH2 as internal trigger signal source for verti...

Страница 8: ...al and horizontal deflection factor of oscilloscope 15 GND Grounding terminal of oscilloscope case 2 Features of rear panel See fig 2 34 AC Power AC Power input jack AC power wire connects here 35 Supporter For supporting the oscilloscope and lead out the power cord Operation methods 3 Basic operation Single channel operation Make sure whether the voltage of the unit is in accordance with your loc...

Страница 9: ...only the signal of trigger signal generator can stably display if TRIG ALT switch depressed both the waveforms stably display 5 Operation By adjusting the setting of Vertical modes switch to add the algebraic sum of CH1 and CH2 will be displayed or algebraic difference if the switch CH2 INV is depressed For gain the accurate values of both the attenuation setting of the both channels must be consi...

Страница 10: ...gger switch set in rising edger triggered and falling edge is triggered for See fig 6 6 Fig 6 Positive zone Level Trigger alternative switch When vertical mode set in dual channel display the switch used in alternative trigger and alternative display suitable for the mode of summarize with CH1 CH2 In alternative mode trigger signal alternates one time for every sweep period This mode is avail for ...

Страница 11: ...on signal adjust compensation capacitor to the best waveform no overshoot fillet upwarping See fig 9 9 Fig 9 Appropriate compensation Overcompensation Undercompensation 11 DC balance adjustment DC BAL Set the input coupling switch of CH1 and CH2 in GND trigger mode in AUTO adjust trace in centre position Adjust the attenuation switch between 5mV and 10mV tune DC BAL until the trace is below zero h...

Страница 12: ...r sampling Trigger switch Z Z axial input High voltage X X axial signal 1 CH1 Sweep generator Line scan switch 2 CH2 Trigger input amplify Trigger signal generator Line scan output amplify External Auto circuit Free oscillate signal To circuits of other units Power circuit 50 60Hz AC 50 60Hz Power trigger 2Vp p 1KHz 2Vp p 1KHz square waveform Calibration signal generator All specifications are sub...

Отзывы: