background image

                (Synchronizing TV-Vertical and TV-Horizontal only in the case of synchronous signals are negative pulses) 

 

Time base 

29) 

  Horizontal sweep speed switch: Sweep speed is divided into 20 steps, from 0.2 S/div to 0.5S/div. And set X-Y 

position can use as X-Y oscilloscope.   

30) 

  Horizontal fine tune: Fine tuning the time of horizontal sweep, in order to calibrate the sweep time in accordance 

with TIME/DIV on panel. The sweep speed of TIME/DIV is continuously changed, when you anticlockwise go 
round and round the base, that is the calibration location. The total delayed time is above 2.5 times.   

32) 

 

  horizontal position: Adjust the horizontal position of trace on screen. 

31) 

  Sweep expand switch: Sweep speed will be magnified 10 times by pressing it. 

 

Others:   

1)

 

  CAL: Provide square signal with 2Vp-p amplitude and 1kHz frequency, which is used in calibrating compensate 

capacitor of 10:1 probe, also detecting the vertical and horizontal deflection factor of oscilloscope. 

15) 

  GND: Grounding terminal of oscilloscope case. 

 

2.

 

Features of rear panel: (See fig.2) 

34) 

  AC Power: AC Power input jack, AC power wire connects here. 

35) 

  Supporter: For supporting the oscilloscope and lead out the power cord. 

 
 
 
 

Operation methods                                                   

3.

 

  Basic operation: Single channel operation 

          Make sure whether the voltage of the unit is in accordance with your local voltage before switch on, then adjust 
the relative control components according to the following form: (Form 4)   
   

(

Functions   

No. 

Setting 

POWER 

6) 

OFF 

INTEN 

2) 

Centralize 

FOCUS 

3) 

Centralize 

VERT MODE 

14) 

CH1 

ALT/CHOP 

12) 

Released(ALT) 

CH2 INV 

16) 

Released 

POSITION 

11) 19) 

Centralize 

VOLTS/DIV 

7) 22) 

0.5V/DIV 

VARIABLE 

9) 21) 

CAL(Calibration position) 

AC-GND-DC 

10) 18) 

GND 

Source 

26) 

TRIG.ALT 

27) 

Released 

TRIGGER MODE 

25) 

Automatic 

TIME/DIV 

29) 

0.5mSee/DIV 

SWP.VER 

30) 

Calibration position 

POSITION 

32) 

Centralize 

X10 MAG 

31) 

Released 

 
After setting the switch and control part, connect power cord, continue: 

 

Switch on, power indicator on , traces appear on screen in about 20s. If no traces appear in 60s, check the settings of 

Содержание AT7328 Series

Страница 1: ...d high light Voltage of accelerated pole up to 2kV it can be able to display clear traces even under high speed sweep The alternative triggering function can observe different signal waveform in two frequencies TV signal synchronous function CH1 output 50 output signal in the rear panel can drive frequency counter directly or other instrument Z axial input Function of bright adjustable can add fre...

Страница 2: ...ce Transformer Coaxial Directional Coupler Detector Power Divider 7 Advanced Milli volt Meter 8 Regulated DC Power Supply 1 AT TPR Series Low Ripple CV Digital Display Power Supply Industrial Products 2 AT E Series Ultra Ripple High Precision CV CC Power Supply Research Education Military Products 3 AT SPS Series Compensation Linear Power Supply 4 AT KPS Series Switching Adjustable Power Supply 5 ...

Страница 3: ...DD Addition of Channel 1 and Channel 2 Interruption Repetitive Frequency Approx 250KHz Input Coupling AC GND DC Max Input Voltage 300Vp p AC Frequency 1KHz When probe sets in 1 1 maximum effective readout value is 40Vpp 14Vrms Sine wave When probe sets in 10 1 maximum effective readout value is 400Vpp 140Vrms Sine wave CMRR 50 1 50KHz sine wave If CH1 and CH2 have the same sensibility 1000 1 50KHz...

Страница 4: ...C AC peak AC not exceed 1KHz Sweep time 0 2 Sec0 5Sec DIV Accuracy 3 Trimming 1 2 5 panel graduation Sweep expansion 10 times 10MAG Sweep time accuracy 5 20n sec 50n Linearity 3 10MAG 5 No calibration at 20ns 50ns Horizontal System Position by 10MAG 2DIV in center CRT RF MICROWAVE INSTRUMENT SERIES Specifications Sensibility Refer to Vertical axis Bandwidth DC 500KHz X Y Mode X Y Phase difference ...

Страница 5: ...g factory open the package immediately as this item is received to check whether there is any spoilage during transportation if yes contact supplier or sender immediately 2 Check power voltage AC 110V 220V is available check the voltage switch whether conforms with the local supply or not before switch on Notice Failure in select power will result in damage the oscilloscope Warning To prevent elec...

Страница 6: ...V peak Warning To prevent damage do not exceed these values Frequency of Max input voltage must be lower than 1 kHz If an AC volt overlaps on DC volt the maximum voltage of CH1 and CH2 input cannot exceed 300V therefore for a AC which average voltage is zero its peak peak value is 600Vpp Operation methods 1 Features of front panel See fig 1 CRT 6 Power Main power switch LED 5 lights when this swit...

Страница 7: ...l and trigger signal of CH2 invert simultaneously Triggering 24 External trigger input Used for trigger for external signal For this case Switch 23 it should be set in EXT 23 Selection of trigger source Trigger of internal INT or external EXT CH1 Choose CH1 as internal trigger signal source for vertical mode switch 14 is set in DUAL or ADD CH2 Choose CH2 as internal trigger signal source for verti...

Страница 8: ...al and horizontal deflection factor of oscilloscope 15 GND Grounding terminal of oscilloscope case 2 Features of rear panel See fig 2 34 AC Power AC Power input jack AC power wire connects here 35 Supporter For supporting the oscilloscope and lead out the power cord Operation methods 3 Basic operation Single channel operation Make sure whether the voltage of the unit is in accordance with your loc...

Страница 9: ...only the signal of trigger signal generator can stably display if TRIG ALT switch depressed both the waveforms stably display 5 Operation By adjusting the setting of Vertical modes switch to add the algebraic sum of CH1 and CH2 will be displayed or algebraic difference if the switch CH2 INV is depressed For gain the accurate values of both the attenuation setting of the both channels must be consi...

Страница 10: ...gger switch set in rising edger triggered and falling edge is triggered for See fig 6 6 Fig 6 Positive zone Level Trigger alternative switch When vertical mode set in dual channel display the switch used in alternative trigger and alternative display suitable for the mode of summarize with CH1 CH2 In alternative mode trigger signal alternates one time for every sweep period This mode is avail for ...

Страница 11: ...on signal adjust compensation capacitor to the best waveform no overshoot fillet upwarping See fig 9 9 Fig 9 Appropriate compensation Overcompensation Undercompensation 11 DC balance adjustment DC BAL Set the input coupling switch of CH1 and CH2 in GND trigger mode in AUTO adjust trace in centre position Adjust the attenuation switch between 5mV and 10mV tune DC BAL until the trace is below zero h...

Страница 12: ...r sampling Trigger switch Z Z axial input High voltage X X axial signal 1 CH1 Sweep generator Line scan switch 2 CH2 Trigger input amplify Trigger signal generator Line scan output amplify External Auto circuit Free oscillate signal To circuits of other units Power circuit 50 60Hz AC 50 60Hz Power trigger 2Vp p 1KHz 2Vp p 1KHz square waveform Calibration signal generator All specifications are sub...