![ATS PAV4 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/ats/pav4/pav4_instruction-manual_3005219038.webp)
MAN - PAV4 - REV: Original | I. Maintenance
33
I. Maintenance
I.1 Preventative Maintenance
1. Keep the pressure vessel lid bolts lubricated using an anti-seize agent. Repeated insertion and extraction of
stainless steel bolts in a stainless steel vessel without using anti-seize paste can cause galling. Lower speed insertion
and extraction of bolts can also help to prevent galling. Use ¼ to ½ the speed for steel in steel.
2. Periodically lubricate the O-ring with silicone lubricant. The O-ring is made of synthetic rubber and is located in the
pressure vessel flange. Damaged or worn O-rings should be replaced with the same material.
I.2 Changing Fuses
1. Control Circuit Fuse (FU1 & FU2): The control circuit fuse is a 2 amp, 250 volt 1/4” x 1-1/4” BUssmann MDC (or
equivalent) fuse.
2. Heater Circuit Fuse (FU3 & FU4): The heater circuit fuse is a 10 amp, 250 volt BUssman GBB (or equivalent) fuse.
I.3 Spare Parts
Some parts are supplied from outside ATS. Refer to manufacturer’s literature.
Consumables
Part Number
Description
PAV1045
PAV O-Ring
PAV-PAN-KIT
Set Of (10) Specimen Pans
PAV-LUBE
Anti-Seize Lubricant
PAV1038
Fuse, 2A 1/4” - 1 1/4” Slow Blow F2
102674
Fuse, 10 Amp 250 Volt
PAV1068
Led, Tri-Color Diffused
PAV1070
Led, Red Diffused 4
PAV1072