9
The manual, which accompanies this product, may be
revised from time to time and must therefore be reviewed
prior to each procedure in which the product is used.
Intended use
The Provox Flush is intended to be used to flush drinking
water or air through the inner lumen of a Provox voice
prosthesis for cleaning purposes. The Provox Flush is
intended for both home and clinical use by patient or
clinician.
WARNINGS
The flush is only recommended for patients with sufficient
manual dexterity acceptable vision and satisfactory
cognitive ability.
Cleaning and disinfection
Clean the Provox Flush after each use according to fig. 9-11.
Do not use any water other than drinking water to clean
and rinse the device.
Disinfect the Provox Flush at least once a day (fig. 12) with
one of the following methods:
• Ethanol 70% for 10 minutes
• Isopropylalcohol 70% for 10 minutes
• Hydrogenperoxide 3% for 60 minutes
Caution:
Do not clean or disinfect by any other method as
this might cause product damage and injury.
If the Provox Flush looks dirty or has air dried in an area
with a risk of contamination the device should be both
English