23
Performance test
The performance test of the unit shall be carried out with the supply voltage given in the operating contract and the given mains
frequency. The unit is operated in the suction mode for this.
The following performance must be checked:
•
Vacuum: > 88%
•
Flow: at least 32 l/min on the unit connection pipes (without bacterial fi lter).
Set delivery condition
The delivery condition must be set as follows:
In Service Mode 1:
•
Adjust unity:
Delivery in Germany: kPa
Delivery abroad: kPa
•
LCD brightness:
20
•
Vacuum limit
50 kPa
•
Increase/decrease time
10 seconds
•
Vacuum hold time
60 seconds
•
Pause
60 seconds
In Service Mode 2:
•
Clear Errors: Yes (erase error memory)
Operation on the unit before switching it off:
•
Vacuum:
Maximum
•
Type of operation:
Suction Mode with Auto-Standby
6.0 Maintenance and repair