![Atmos MEDAP FINA DFLOW Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/atmos/medap-fina-dflow/medap-fina-dflow_operating-instructions-manual_3003646019.webp)
Operation
Function test
GA 5752 2804 GB 18
19
4
4
Operation
4.1
Function test
DANGER!
Function check!
The product is used in the treatment of patients. Any restriction in the unit's
performance can result in serious complications in treatment.
Perform a complete function check every time before using the unit.
Perform a complete function check of the tapping unit prior to use. During functional testing, pay
attention to the following conditions:
Version A: Tapping unit with integrated gas pin
• The tapping unit is correctly plugged into the terminal unit.
Version B: Tapping unit with rail clamp and NIST connection
• The gas probe of the connection tube is correctly inserted in the terminal unit.
• The NIST nipple of the connection tube is correctly secured in the NIST connection of the
tapping unit.
• The NIST screw connection is tightly fastened.
• The tapping unit is locked firmly to the equipment rail.
All versions:
• The tube connectors are firmly secured and tightly sealed, and no mechanical forces are act-
ing on the tubes.
• The plastic and rubber components are in perfect condition and show no signs of ageing.
• The accessories are correctly connected.
• The product is leak-free.
• It is possible to variably adjust the flow rate from zero to the maximum flow.
• The product is in good hygienic condition.
4.2
Setting the flow for treatment
DANGER!
Over-pressure!
The product is under pressure whenever it is connected to the terminal unit, even if
the valve is closed.
The flowmeter viewing tube may not be unscrewed when the product is connected.
WARNING!
Compressed gas setting!
The regulating mechanism is sensitive. Make the compressed gas settings very
carefully!
NOTE
The rate of gas flow is displayed on the flowmeter viewing tube in the centre of the
indicator ball.
Содержание MEDAP FINA DFLOW
Страница 1: ...OPERATING INSTRUCTIONS FLOWMETER MEDAP FINA FLOW MEDAP FINA DFLOW GA 5752 2804 GB 18 ...