1.0 General Notes
1.4 For your safety
•
The ATMOS S 61 Servant ENT workstation is produced
according to IEC 601/ EN 60601 and listed in the
following classes:
•
VDE Class of protection 1
•
Class IIa (EEC 93/42).
•
The device may only be connected to a properly installed
shockproof socket. Correct confi guration for installation
to country-specifi c connections:
green/yellow: protective conductor (PE)
blue: neutral conductor (N)
black resp. brown: phase (L)
•
Do not constrict the air supply on the back of the unit!
•
The ATMOS S 61 Servant may be used in supervised ope-
ration by qualifi ed personnel only which has been authori-
sed by ATMOS and which has been trained for operating
the appliance (MPBetreibV §2 Abs. 1 and 2).
•
The mains voltage specifi ed on the type plate must corre-
spond with the data of the power supply system.
•
Prior to opening, switch off the unit and separate it from
the line voltage.
•
The operation via an isolating transformer is only permitted
when an insulation monitoring or another type of monitoring
function for electrical protection is provided.
•
The ambient conditions specifi ed in section „Technical
specifi cations“ must be strictly observed!
•
Switch off main switch after fi nishing work in practice and
close water supply, if present.
•
Attention when working with endoscopes near the light
sources. Do not look directly into the cone of light! At a
contingent light blackout slowly remove the endoscope
out of the working area!
•
The ATMOS S 61 Servant may be operated only in rooms
used for medical purposes, but not in areas subject to
explosion hazards.
•
The ATMOS S 61 Servant fully complies with the electro-
magnetic immunity requirements of standard IEC 601-1-2/
EN 60601-1-2 „Electromagnetic compatibility - Medical
Electrical Equipment“.
•
The ATMOS S 61 Servant may not be operated with units
not complying with the requirements of standard EN
60601-1 „Medical Electrical Equipment“ and EN 60601-
1-2 „Electromagnetic compatibility (Medical Electrical
Equipment)“.
•
ATMOS cannot be hold liable personal injury and damage
to property, if
•
no original ATMOS parts are used,
•
the instructions for use in the operating instructions or
in these service instructions are not being observed,
•
assembly, new settings, alterations, extensions and
repairs have been carried out by personnel not autho-
rised by ATMOS.
•
These service instructions correspond with the construction
of the unit and with the current status of safety-related
standards at the time of printing. Proprietary rights are
existing for all described circuits, processes, names, soft-
ware programs and units.
•
During function tests at the open device all measures
must be taken which are required for protection against
an electric shock.
•
Attention! Danger of injury at rotating pumps respectively
motors.
•
Attention! When working at the light module, there is a
danger of eye injury (dazzlement or burn) caused by the
extreme brightness. In addition there is the danger of skin
burn due to the extremely high temperature generated
by the bulb.
•
During service work at the ATMOS S 61 Servant - espe-
cially at the suction system - hygiene regulations must
strictly be observed.
•
After each maintenance or service work a function test
must be performed according to the check list as well
as a safety-related inspection acc. to DIN EN 0751-1.
ATMOS recommends an inspection according to
the testing instructions and a safety-related in-
spection once a year.