ATI
Q45P pH System
Part 5
–Calibration
33
O&M Manual
Rev-B (7/15)
buffers are not present in this menu, the user can override the automatic values
and enter arbitrary values. However, the highest accuracy is provided when the
user selects and uses buffers from this pre-defined table list. With the pre-
defined buffers, the temperature variations in the buffer are automatically
compensated for during the calibration process. If the buffer data is manually
entered, the calibration buffer sample must be very temperature stable to achieve
the same degree of accuracy.
Procedure
1.
Remove sensor from application. Rinse and clean if necessary.
2.
Allow sensor to temperature equilibrate with the buffer as best as possible. With
the sensor coming from an application solution that differs greatly in temperature
from the buffer, the user may have to wait as much as 20 minutes for this to
occur.
3.
Scroll to the CAL menu section using the MENU key and press ENTER or the UP
arrow key.
Cal pH
will then be displayed.
4.
Press the ENTER key. The screen will display a flashing 1 for 1-point or a 2 for
2-point calibration. Using the UP arrow key, set for a 2-point calibration and
press ENTER.
5.
The display will prompt the user to place the sensor in the first buffer and press
ENTER. If the sensor has been placed into this buffer already, once the
temperature has stabilized, press ENTER to continue.
6.
The present pH value will be displayed and the secondary line of the display will
flash
Wait
for approximately 10-15 seconds. At this time the system is
attempting to recognize the first buffer value from the two values entered into the
Set Buffers selection.
7.
The screen will display the buffer value to be used for calibration. If the user
chooses to change this value, the arrow keys can be used to modify the value.
Any value between 0.00 and 14.00 pH can be entered. After adjusting this value,
or to accept the automatic value, press ENTER.
8.
The system now begins acquiring data for the calibration value of this buffer
point. As data is gathered, the units for pH and temperature may begin to flash.
Flashing units indicates that this parameter is unstable. The data point