ATI
Q45P pH System
Part 3
–Installation
11
O&M Manual
Rev-B (7/15)
3.3
Combination Electrode Connection
The Q45P may also be used with non-amplified simple combination electrodes
(see Figure 4). Note that a wire jumper must be installed from Terminal 3 to
Terminal 8. The user must also select Sensor Type
2
within the Config Menu
(see Section 4.9). The maximum sensor-to-instrument cable length will be
severely limited (30-50 feet) with electrodes of this type. The length will depend
on the specific electrode impedance and the quality of interconnect cable
provided by the manufacturer.