Chapter 3 –
Start up, adjust and measures
UM-23900B-U User
manual
ATEQ CDF60
Page 13/40
3. PARAMETERS ADJUST
It's existing three adjusted parameters.
3.1. F
ILTER
It can be adjusted from 0.1 to 9.9 seconds by 0.1 second of step. It allows making a
mean on the measurement time, making easier the measurement reading.
Enter into the parameters menu, by pressing and
old the
ENTER
during 3 seconds.
+ / - 4 0 . 0 0 c m 3 / m i n
A I R S T D . C O N D .
FLOW :
09.36
TEMP. :
25.8°C
ATM.P :
1010.5 hPa
The
PARAMETER
menu is displayed. By using the
UP
and
DOWN
key select
FILTER
parameter.
PARAMETER
>FILTER : 0.5 s
FLOW UNIT.: cm3/min
STD.COND.: YES
Then press the
ENTER
key to modify the
parameter (the cursor moves on the right).
PARAMETER
FILTER : 0.5 s <
FLOW UNIT.: cm3/min
STD.COND.: YES
By using the
UP
and
DOWN
key
modify the parameter to the hoped value.
PARAMETER
FILTER : 1.0 s <
FLOW UNIT.: cm3/min
STD.COND.: YES
Validate with the
ENTER
key.
PARAMETER
>FILTER : 1.0 s
FLOW UNIT.: cm3/min
STD.COND.: YES
To return to the measurement mode, press
CANCEL
.
+ / - 4 0 . 0 0 c m 3 / m i n
A I R S T D . C O N D .
FLOW :
09.36
TEMP. :
25.8°C
ATM.P :
1010.5 hPa
Содержание CDF60
Страница 1: ...USER MANUAL ATEQ CDF60 Version 1 0 Reference UM 23900B U...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 20: ...Chapter 2 User Interfaces UM 23900B U User manual ATEQ CDF60 Page 10 40...
Страница 40: ...Chapter 6 Error messages UM 23900B U User manual ATEQ CDF60 Page 30 40...
Страница 46: ...Chapter 7 PC installation UM 23900B U User manual ATEQ CDF60 Page 36 40...
Страница 50: ...Index UM 23900B U User manual ATEQ CDF60 Page 40 40...
Страница 51: ......