Chapter 3 –
Start up, adjust and measures
UM-23900B-U User
manual
ATEQ CDF60
Page 15/40
3.3. S
TANDARD CONDITIONS
(S
TD COND
)
This parameter allows displaying the measurements values corrected to the standards
conditions values, atmospheric pressure and temperature, to the user preferences.
Enter into the parameters menu, by pressing and
old the
ENTER
during 3 seconds.
+ / - 4 0 . 0 0 c m 3 / m i n
A I R
FLOW :
09.36
TEMP. :
25.8°C
ATM.P :
1010.5 hPa
The
PARAMETER
menu is displayed. By using the
UP
and
DOWN
key select
STAND. COND
parameter.
PARAMETER
FILTER : 1.0 s
FLOW UNIT.: cm3/min
>STD.COND.: NO
Then press the
ENTER
key to modify the
parameter (the cursor moves on the right).
PARAMETER
FILTER : 1.0 s
FLOW UNIT.: cm3/min
STD.COND.: NO <
By using the
UP
and
DOWN
key
validate the function by
YES
. Validate with the
ENTER
key.
PARAMETER
FILTER : 1.0 s
FLOW UNIT.: cm3/min
STD.COND.: YES <
The standards temperature and atmospheric
pressure values are displayed, it's possible to
modify following the user preferences. Default
values: 20° C and 1013.2 hPa.
STANDARD CONDITION
>TEMP. : 20.0°C
PRESS. : 1013.2hPa
To return to the measurement mode, press twice
CANCEL
. The measurements values
displayed are corrected following standards
conditions.
+ / - 4 0 . 0 0 c m 3 / m i n
A I R S T D . C O N D .
FLOW :
09.36
TEMP. :
25.8°C
ATM.P :
1010.5 hPa
Содержание CDF60
Страница 1: ...USER MANUAL ATEQ CDF60 Version 1 0 Reference UM 23900B U...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 20: ...Chapter 2 User Interfaces UM 23900B U User manual ATEQ CDF60 Page 10 40...
Страница 40: ...Chapter 6 Error messages UM 23900B U User manual ATEQ CDF60 Page 30 40...
Страница 46: ...Chapter 7 PC installation UM 23900B U User manual ATEQ CDF60 Page 36 40...
Страница 50: ...Index UM 23900B U User manual ATEQ CDF60 Page 40 40...
Страница 51: ......