[22]
Communication parameters of ALP-‐94 RS-‐232 are as following. There are no settings in the instrument to
change these parameters.
Communication is one-‐way from the instrument to the receiver.
Data’s format :
RS232 EOL (16bit data + EOL)
Baud rate
:
57600bps
Data bits
:
8
Stop bits
:
1
Parity
:
none
Over
.
“
999999” message is displayed on the screen, when input value is too high.
“
SCL-HI
” value can be increased (see 4.3.1. Setting of Scale Limit Values p. 16).
under
.
“
999999” message is displayed on the screen, when input value is too small.
“
SCL-Lo
” value can be decreased (see 4.3.1. Setting of Scale Limit Values p. 16).
adcerr
There is an error at the analog input section of the instrument. The
measurement cycle cannot be completed. Product should be given to authorized service.
5.
DATA
PROTOCOL
6.
ERROR
MESSAGES