Sonelastic
®
Systems
SA-BC Adjustable support for bars and cylinders
Pág. 22 de 24
11. Warnings and support transportation
Reading all the information contained in this Installation and Operation Manual is
compulsory for the correct use of the support;
The power outlet where the computer will be connected must have a functional ground
pin;
Do not use this support for other purposes apart from the ones specified by this manual;
The non-compliance with the instructions provided by this manual in what regards the use
of the support may reduce or invalidate warranty time.
Support transportation:
- Transport the support with care;
- Avoid impacts and falls when transporting the support;
- Do not transport the support under rain or snow, even when wrapped in its original
packaging.
12. Maintenance and troubleshooting
- Depending on the specimen material, detachment of residues may occur during handling.
To avoid hazards and possible damages to the support, clean frequently using a slightly damp
cloth.
- To maintain the support in a good condition and extend its life, keep all accessories, optional
items and chutes clean.
Troubleshooting:
Problem
Possible Cause
Solution
Locked (unable to move)
tightener and
̸
or
anchoring terminal.
The terminals sliding
chutes may be dirty or
blocked.
Vacuum the chutes and remove any
object or residue that may be
obstructing it.
The support wobbles when
placed over a bench.
The bench surface may not
be completely flat or may
present some
irregularities.
Adjust the rubber feet until they are all
well placed and flat on the surface.
The specimen touches the
safety cables and it
disturbs the
measurements.
Specimen weight is above
the limit or the Support-
cable is not under enough
tension.
Check the weight limits in
6
–
Specifications
. Check the correct way to
apply tension to the support-cables in
8.4
–
Replacing the support-cable set.
The specimen does not fit
between the cables
supporting tubes.
Specimen dimensions are
above the limit.
Check dimensions limits in
6
–
Specifications.