P-OQ1011-3276-J01
9
Operational
Qualification (OQ)
Automatic Digital
Refractometer RX-5000i
Operational and Performance Qualification Test Results Report
Results
Notes
Test date
Model name
Serial number
Printer name
Serial number
Ambient
Condition
Temperature
℃
Relative Humidity
%
1
.
Performance Test Results
(20
℃
)
Distilled water
(1.33299 nD)
10.00% Brix
standard
sucrose
solution
30.00% Brix
standard
sucrose
solution
1.51 nD
standard liquid
1
st
measurement
2
nd
measurement
3
rd
measurement
4
th
measurement
Average
Margin of error
Standard of evaluation = ±0.05% Brix, ±0.00006 nD accuracy
2
.
Function Test Results
Items
Results
Switch
□
Power switch
Screens and icons
□
Home screen
□
Set up screen
□
ZERO icon
□
START icon
Display
□
Opening message
□
Date and time
□
Display
Messages
□
ZERO SET END
□
OUT OF WATER
□
OUT OF RANGE
□
OUT OF SCALE
Output
*1
□
Printer
□
USB flash drive
RS-232C
*1
□
Data output
□
Operation
*1 Test items when connecting to a printer or PC.
3
.
Final Evaluation
Evaluate the performance and function test results and make the verdict.
□
Pass
□
No-Pass
Test Date:
Approval Date: