P-OQ1011-3276-J01
8
Operational
Qualification (OQ)
Automatic Digital
Refractometer RX-5000i
manual.
7) Type in the start command “*START” and press “ENTER” on the PC. A measurement
will be taken.
3-6. USB Data Storage Test
1) Connect an USB flash drive with enough space to the instrument.
2) Touch “DATA” on the home screen.
3) The past measurement data will be recalled. Touch “RESULT” on the bottom of the
screen.
4) The file name screen will appear. Type in the file name to be saved on the USB drive
and touch “ENTER.”
5) “SAVING” will be displayed on the screen. The data have successfully been copied to
the USB drive once “COMPLETED” is displayed.
6)
Disconnect the USB drive from the instrument and connect it to a PC to check the
contents.
4
.
Publication to Operational and Performance Qualification Test Results
Report
Record the function test items on the Operational and Performance Qualification Test
Results Report.
Evaluating Performance of RX-5000i Automatic Digital Refractometer
Compare and evaluate the performance and function test results recorded on the
Operational and Performance Qualification Test Results Report against the standard of
evaluation. Have the operator and an approver sign the report.