![ATAGO PRM-2000 A Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/atago/prm-2000-a/prm-2000-a_instruction-manual_2995841039.webp)
38
Examples of entry
If you enter concentrations as 0.0%, 10.0%, 20.0%, enter 1 for the number of display digits of a
concentration.
If you enter concentrations as 10.000%, 20.000%, 30.000%, enter 3 for the number of display digits of
a concentration.
If you enter concentrations as 0.20000, 0.40000, 1.00000, 1.50000, enter 5 for the number of display
digits of a concentration.
m1
1
A displayable value is maximum seven characters including the decimal point.
e.g. 1.00000 10.0000 100.000 1000.00 10000.0
m1
2
Resolution to the recorder output is 30000 data.
If above, the unit cannot output correctly.e.g., in case of 0.000-100.000, limit the range of recorder
output to 0.000-30.000 or 20.000-50.000, or reduce the digit number such as 0.00-100.00 to make the
data below 30000.
Example
0.000
~
30.000 : 30000 data
0.000
~
50.000 : 50000 data (above )
1.0000
~
5.0000 : 40000 data (above )
1.000
~
5.000
: 4000 data
m1
3
Enter the alphabet in uppercase. (See your computer's manual)
The * mark is an asterisk (ten-key's multiplication symbol).
Fig.11-18
⑦
Once the data entry is complete (Fig.11-18), save the file.
fig. 11-19