27
⑧
Press the ENTER key to access the SET UP mode.
The most recent measurement value will be displayed.
⑨
Use
←
→
key to move digit position and then use
↑
↓
key to set a standard value (Fig.11-4 and Fig.11-5).
⑩
After setting a standard value, press the ENTER key to
start data writing.
The measurement value display turns off during the data
writing process. This may take a while.
The bias is set and calibration is complete.
The "1. bIAS" screen will be displayed (Fig.11-3).
⑪
To abort while in the SET UP mode, press the SCALE
key.
"StoP" will appear after 10 seconds of inactivity
(Fig.11-2).To start taking measurements with the
post-calibration data press the START/STOP key.
Fig.11-4
Example using the Refractive Index display
Fig.11-5
Fig.11-6
If the last measurement value was LLL.LLL or
HHH.HHH, the screen for setting the bias will not
appear.
Press the SCALE key to return to the previous "1. bIAS"
screen (Fig.11-3).
Press the SCALE key to abort the change and return to
the previous "1. bIAS" screen (Fig.11-3). Press
START/STOP key to reset the factory default setting.
"CL" appears on the measurement value screen
(Fig.11-6), and it returns to the previous "1. bIAS"
screen.