33
(2)
Creating Data Table
(2-1) Definition and Format of Data Table
①
The data table consist of the following three definitions: temperature, Refractive Index, and
Concentration.
②
The maximum number of data points possible is: 6 for temperature and 7 for C oncentration.
③
The temperature is displayed by 3 digits and 1 decimal place (000.0).
The Concentration is expressed by a 6-digit number.
A decimal point can be placed anywhere within the 6 digits.
The Refractive Index (nD) is expressed by a 5-digit number.
In case of a 4-digit number, add a "0 (zero)" at the end to make it 5 digits long.
④
Create a table based on the information above.
Table 11-4
C1
C2
C3
C4
C5
C6
C7
T1
nD10
nD11
nD12
nD13
nD14
nD15
nD16
T2
nD20
nD21
nD22
nD23
nD24
nD25
nD26
T3
nD30
nD31
nD32
nD33
nD34
nD35
nD36
T4
nD40
nD41
nD42
nD43
nD44
nD45
nD46
T5
nD50
nD51
nD52
nD53
nD54
nD55
nD56
T6
nD60
nD61
nD62
nD63
nD64
nD65
nD66
※
C1
~
C7 : Concentration or dilution coefficient, etc.
Enter up to 7 points. Fill in any blank cells with "0.0".
Input condition :
C1
<
C2
<
C3
<
C4
<
C5
<
C6
<
C7
C1
>
C2
>
C3
>
C4
>
C5
>
C6
>
C7
※
T1~T6 : Measurement temperature
Enter up to 6 points. Fill in blank cells, if any, with "0.0".
Input condition : T1
<
T2
<
T3
<
T4
<
T5
<
T6
※
nD10~nD66 : Refractive Index (nD) by Concentration and by temperature
Input condition : nD10
<
nD11
<
nD12
<
nD13
<
nD14
<
nD15
<
nD16
nD20
<
nD21
<
nD22
<
nD23
<
nD24
<
nD25
<
nD26
nD30
<
nD31
<
nD32
<
nD33
<
nD34
<
nD35
<
nD36
nD40
<
nD41
<
nD42
<
nD43
<
nD44
<
nD45
<
nD46
nD50
<
nD51
<
nD52
<
nD53
<
nD54
<
nD55
<
nD56
nD60
<
nD61
<
nD62
<
nD63
<
nD64
<
nD65
<
nD66
※
Fill in any blank cells with "0.0."
※
If the measured value is outside the range in this table, "LLLLLL" or "HHHHHH" is displayed.
※
If the measured temperature is outside the range in this table, "d. Err" is displayed.