Operational 

Q u alification  (O Q

 

 

 

Automatic Polarimeter 

AP-300 

 

Operation Qualification Results Report 

 

     Items 

Descriptive contents 

  Remarks   

Date of test   

 

 

Product name 

and model name 

 

 

Serial number 

 

 

Printer name 

 

 

Serial number 

 

 

Environmental 

conditions 

Temperature               °C 

Relative humidity          % 

 

 

1. Performance Test Data 

 

Standard value 

Standard 

quartz 

control plate (34.77°) 

standard sucrose 

solution of 100°Z 

1st test 

 

 

2nd test 

 

 

3rd test 

 

 

4th test 

 

 

Average value 

 

 

Difference 

 

 

           Judgment standard   Accuracy    Angle of rotation ±0.03 ° 

 

2. Function Examination 

Functions 

Contents 

Switch function 

□ Power supply switch

 

Key functions 

□ZERO          □QUIT          □MENU          □SCALE

 

□START        □ENTER      □Arrow keys

 

□NUMERIC keypad

 

Display functions 

□ Opening message

 

□ Date and time

 

Error messages 

□ ZERO SET END      □OUT OF RANGE

 

□ TEMP. ERROR

 

Output functions 

□ Printer output    □RS

-232C output 

*check each box 

□.

 

 

3. Evaluation 

The results of the performance test and a functional examination were evaluated. 

□    Pass                                          □    Fail

 

 

Date of operation: Year     Month     Day       

Date of approval: Year       Month     Day 

Operator signature: ___________________   

Supervisor signature: ____________________ 

 

Содержание AP-300

Страница 1: ...tionary measurement standard Establish a set measurement standard for your samples taking the accuracy of the AP 300 into account Please refer to the product literature and or the instruction manual for the specifications features and measurement accuracy of the AP 300 These specifications are guaranteed at the time the AP 300 is shipped as a brand new product It is difficult to obtain an accuracy...

Страница 2: ...ailable prepare a standard sucrose solution of 100 Z Refer to below in a 200mm observation tube 3 Recording chart for measurement data How to make standard sucrose solution of 100 Z Put 26 000g of pure sucrose in a graduated cylinder and add distilled water to the pure sucrose until the total volume is 100ml 5 Preparation of Instrument 1 Turn on the power of the AP 300 The measurement screen appea...

Страница 3: ...Check if they are consistent 2 The AP 300 stores the measurement value of the last 30 measurements in the memory If the number of measurements exceeds 30 the newest measurement values will replace the oldest measurement values 3 The measurement values stored in the memory can be recalled on the screen to compare with those in the measurement data recording chart 4 The measurement values stored in ...

Страница 4: ...z control plate 34 Part number RE 72044 Quartz control plate 17 Part number RE 72043 Quartz control plate 8 Use a Quartz control plate whose angle is close to the angle of rotation usually measured 2 If a quartz control plate is not available fill a 200mm observation tube with a prepare a standard sucrose solution of 100 Z 3 Function Examination After a performance test in MODE 1 the measurement s...

Страница 5: ...key 5 The display returns to the SET UP MENU screen Press the QUIT key The display returns to the measurement screen 6 Set a standard Quartz control plate or a standard sucrose solution of 100 Z in the sample chamber 7 Close the sample chamber cover Press the START key 8 OUT OF RANGE will be displayed after measurement is completed TEMP ERROR 1 Press the SCALE key in the measurement screen to disp...

Страница 6: ...ress the MENU key in the measurement screen to display SET UP MENU 2 2 Select 7 SYSTEM LEVEL and press the ENTER key The SYSTEM LEVEL screen appears Make sure the SYSTEM LEVEL is 0 Press the QUIT key The screen returns to SET UP MENU 2 2 Press the QUIT key again to return to the measurement screen ...

Страница 7: ...st Average value Difference Judgment standard Accuracy Angle of rotation 0 03 2 Function Examination Functions Contents Switch function Power supply switch Key functions ZERO QUIT MENU SCALE START ENTER Arrow keys NUMERIC keypad Display functions Opening message Date and time Error messages ZERO SET END OUT OF RANGE TEMP ERROR Output functions Printer output RS 232C output check each box 3 Evaluat...

Отзывы: