Operational 

Q u alification  (O Q

 

 

 

Automatic Polarimeter 

AP-300 

explained later in the manual. 

Press the QUIT  key  in the “SET UP MENU 2/ 2”  screen. The measurement screen  is 

displayed.  Press the SCALE key to change the display. Press the SCALE key until the 
angle of rotation screen appears

 

 

3-2. Error Messages 

 

“ZERO SET ERROR”   

1)

 

Set the standard  quartz control plate  or the standard sucrose solution  of  100°Z in the 

sample chamber. 

2)

 

Close the sample chamber cover. Press the ZERO key. 

3)

 

“ZERO SET Reading” will be displayed. 

4)

 

“ZERO SET ERROR will be displayed. 

Zero-setting can be performed within the range of –5° and + 5° of Angle of Rotation. 

When it is outside of this range, “ZERO SET ERROR” will be displayed. 

 

 

“OUT OF RANGE” 

1)

 

Set the Upper and Lower Limit Bars. 

2)

 

Select “4. LIMIT BAR” from “SET UP MENU 1/ 2” and press the ENTER key. 

3)

 

“LIMIT BAR” is displayed. Enter appropriate values and the name of the sample in “No. 

SAMPLE NAME.” Select “ANGLE OF ROTATION” for “SCALE.”   

4)

 

For the item “LIMIT,” enter “10.00” in HIGH [    ] and “0.00” in LOW [   ]. Press the 

ENTER key. 

5)

 

The display returns to  the  “SET UP MENU”  screen. Press the QUIT key. The display 

returns to the measurement screen. 

6)

 

Set a standard Quartz control plate or a standard sucrose solution of 100°Z in the sample 

chamber.   

7)

 

Close the sample chamber cover. Press the START key. 

8)

 

“OUT OF RANGE” will be displayed after measurement is completed.   

 

“TEMP. ERROR” 

1)

 

Press the SCALE key in the measurement screen to display the International Sugar Scale 

TC (Automatic Temperature Compensation) screen. 

2)

 

Remove the thermal sensor from the sample chamber. 

3)

 

Set a standard Quartz control plate or a standard sucrose solution of 100°Z in the sample 

chamber. 

4)

 

Close the sample chamber cover and press the START key. 

5)

 

“TEMP. ERROR” will be displayed after approximately 45 seconds.   

6)

 

Re-install the thermal sensor. 

 

3-3. Password Function 

  The password “2222” and the system level “0” are the default  factory settings of the 

password function. Use this default password to change the password for the first time. 

 

  The password and the system level are important security settings. This chapter only 

explains how to check the initial screens of 7. SYSTEM LEVEL and 8. PASSWORD 

CHANGE. The password function can be changed once the user decides on the password 

and system level. 

 

 

Содержание AP-300

Страница 1: ...tionary measurement standard Establish a set measurement standard for your samples taking the accuracy of the AP 300 into account Please refer to the product literature and or the instruction manual for the specifications features and measurement accuracy of the AP 300 These specifications are guaranteed at the time the AP 300 is shipped as a brand new product It is difficult to obtain an accuracy...

Страница 2: ...ailable prepare a standard sucrose solution of 100 Z Refer to below in a 200mm observation tube 3 Recording chart for measurement data How to make standard sucrose solution of 100 Z Put 26 000g of pure sucrose in a graduated cylinder and add distilled water to the pure sucrose until the total volume is 100ml 5 Preparation of Instrument 1 Turn on the power of the AP 300 The measurement screen appea...

Страница 3: ...Check if they are consistent 2 The AP 300 stores the measurement value of the last 30 measurements in the memory If the number of measurements exceeds 30 the newest measurement values will replace the oldest measurement values 3 The measurement values stored in the memory can be recalled on the screen to compare with those in the measurement data recording chart 4 The measurement values stored in ...

Страница 4: ...z control plate 34 Part number RE 72044 Quartz control plate 17 Part number RE 72043 Quartz control plate 8 Use a Quartz control plate whose angle is close to the angle of rotation usually measured 2 If a quartz control plate is not available fill a 200mm observation tube with a prepare a standard sucrose solution of 100 Z 3 Function Examination After a performance test in MODE 1 the measurement s...

Страница 5: ...key 5 The display returns to the SET UP MENU screen Press the QUIT key The display returns to the measurement screen 6 Set a standard Quartz control plate or a standard sucrose solution of 100 Z in the sample chamber 7 Close the sample chamber cover Press the START key 8 OUT OF RANGE will be displayed after measurement is completed TEMP ERROR 1 Press the SCALE key in the measurement screen to disp...

Страница 6: ...ress the MENU key in the measurement screen to display SET UP MENU 2 2 Select 7 SYSTEM LEVEL and press the ENTER key The SYSTEM LEVEL screen appears Make sure the SYSTEM LEVEL is 0 Press the QUIT key The screen returns to SET UP MENU 2 2 Press the QUIT key again to return to the measurement screen ...

Страница 7: ...st Average value Difference Judgment standard Accuracy Angle of rotation 0 03 2 Function Examination Functions Contents Switch function Power supply switch Key functions ZERO QUIT MENU SCALE START ENTER Arrow keys NUMERIC keypad Display functions Opening message Date and time Error messages ZERO SET END OUT OF RANGE TEMP ERROR Output functions Printer output RS 232C output check each box 3 Evaluat...

Отзывы: