P-OQ0612-5291-E03 

4/7 

Operational 

Q u alification  (O Q

 

 

 

Automatic Polarimeter 

AP-300 

displayed. It can also export signals to a printer. This chapter explains how to test such functions. 

The function  examination should be performed by  a regular user of the AP-300 who is familiar 

with the operation.  The examination process includes unusual operations, such as displaying 

error messages intentionally. Carefully follow the instructions in this manual. 

 

 

1. Preparation of Instrument 

 

1) If using an exclusive printer, connect it to the AP-300. 

2)

 

Turn on the power of the AP-300. The opening message will be printed out. 

3)

 

Leave the instrument on for at least 30 minutes. 

4)

 

Make sure that “MODE-1” is set. 

5)

 

Check the date and time on the display screen of the AP-300. 

6)

 

Check the settings on the printer.   

Set the date, time, average values, sample numbers, etc. 

Measurement values displayed on the LCD will be printed out. 

 

 

2. Items to Prepare 

 

   

1) Standard quartz control plate

with traceability certificate) 

 

Part number RE-72045 

Quartz control plate 34° 

Part number RE-72044 

Quartz control plate 17° 

Part number RE-72043 

Quartz control plate 8° 

 

 U

se a Quartz control plate whose angle is close to the angle of rotation usually measured.   

 

2)  If a quartz control plate  is not available, fill a 200mm observation tube with a prepare a 

standard sucrose solution of 100°Z   

 

 

 

3. Function Examination 

After a performance test in MODE-1, the measurement screen will be displayed. 

 

3-1. Key Functions in Each Mode 

 

1)

 

Press the MENU key to display "SET UP MENU 1/ 2."   

Press the cursor keys (

↓,↑)

 to check if the pointer moves from the menu items (1) through 

(6).   

Select  a  number,  using  the  cursor  keys  (↓,↑),  and  press  the  ENTER  key  to  enter  the 

selected menu item. Check the display and set the values. 

2)

 

Press the MENU key in the “SET UP MENU 1/ 2” screen to display “SET UP MENU 2/ 2.”   

Press the cursor keys (↓,↑) to check if the pointer moves from the menu items (1) through 

(8).   

Select  a  number,  using  the  cursor  keys  (↓,↑),  and  press  the  ENTER  key  to  enter  the 

selected menu item. Check the display and set the values. Press the ENTER key to return 

to the "SET UP MENU 2/ 2" screen. Check all the menu items (1) through (6) in the same 

manner except for the items (7) and (8). These  are the password functions and will be 

Содержание AP-300

Страница 1: ...tionary measurement standard Establish a set measurement standard for your samples taking the accuracy of the AP 300 into account Please refer to the product literature and or the instruction manual for the specifications features and measurement accuracy of the AP 300 These specifications are guaranteed at the time the AP 300 is shipped as a brand new product It is difficult to obtain an accuracy...

Страница 2: ...ailable prepare a standard sucrose solution of 100 Z Refer to below in a 200mm observation tube 3 Recording chart for measurement data How to make standard sucrose solution of 100 Z Put 26 000g of pure sucrose in a graduated cylinder and add distilled water to the pure sucrose until the total volume is 100ml 5 Preparation of Instrument 1 Turn on the power of the AP 300 The measurement screen appea...

Страница 3: ...Check if they are consistent 2 The AP 300 stores the measurement value of the last 30 measurements in the memory If the number of measurements exceeds 30 the newest measurement values will replace the oldest measurement values 3 The measurement values stored in the memory can be recalled on the screen to compare with those in the measurement data recording chart 4 The measurement values stored in ...

Страница 4: ...z control plate 34 Part number RE 72044 Quartz control plate 17 Part number RE 72043 Quartz control plate 8 Use a Quartz control plate whose angle is close to the angle of rotation usually measured 2 If a quartz control plate is not available fill a 200mm observation tube with a prepare a standard sucrose solution of 100 Z 3 Function Examination After a performance test in MODE 1 the measurement s...

Страница 5: ...key 5 The display returns to the SET UP MENU screen Press the QUIT key The display returns to the measurement screen 6 Set a standard Quartz control plate or a standard sucrose solution of 100 Z in the sample chamber 7 Close the sample chamber cover Press the START key 8 OUT OF RANGE will be displayed after measurement is completed TEMP ERROR 1 Press the SCALE key in the measurement screen to disp...

Страница 6: ...ress the MENU key in the measurement screen to display SET UP MENU 2 2 Select 7 SYSTEM LEVEL and press the ENTER key The SYSTEM LEVEL screen appears Make sure the SYSTEM LEVEL is 0 Press the QUIT key The screen returns to SET UP MENU 2 2 Press the QUIT key again to return to the measurement screen ...

Страница 7: ...st Average value Difference Judgment standard Accuracy Angle of rotation 0 03 2 Function Examination Functions Contents Switch function Power supply switch Key functions ZERO QUIT MENU SCALE START ENTER Arrow keys NUMERIC keypad Display functions Opening message Date and time Error messages ZERO SET END OUT OF RANGE TEMP ERROR Output functions Printer output RS 232C output check each box 3 Evaluat...

Отзывы: