ASTUTE140
®
Meter
ASTUTE140
®
Meter Operation
©Astute Medical, Inc. 2021 PN 300005 Rev G 2021/03/03
63
iii. Run high and low LQC measurements (See “External Liquid Quality Control” on
p. 56 for instructions).
iv. When high and low LQC measurements have passed, repeat steps 1–9.
10. If LQC measurements for the Test cartridge lot have passed, a screen indicating the
amount of time remaining until test completion will be displayed. To abort the test for any
reason, select
Cancel
by pressing the left soft key.
11. Once the test is complete, the test results will be displayed.
12. To manually print the patient test results if Autoprint is disabled:
a. Press the right soft key to display the
Options
pop-up menu.
b. When the
Options
pop-up menu is displayed,
is highlighted. To print the
patient test results, press the right soft key.
c. To hide the
Option
s pop-up menu, use the
key to highlight
Hide Options
and
press the right soft key.
d. To return to the
Main Menu
, press the left soft key.
13. Press the eject button to the left of the LCD screen to open the ASTUTE140
®
Meter drawer.
Remove the Test cartridge and discard according to the laboratory’s standard practices for
the type of biological sample used.
14. To run the next patient sample after viewing or printing the patient test results, press the
right soft key to display the
Options
pop-up menu. Use the
key to highlight
Next
Patient
and press the right soft key. Repeat steps 4–13.
Review and Management of Test Results
Patient, LQC and EQC test results can be recalled, printed or sent to the LIS at any time. To
perform these functions, follow the instructions below.
Patient Results
1. Press the
key to display the
Main Menu
.
Содержание ASTUTE140
Страница 1: ...Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 1 ASTUTE140 Meter User Manual...
Страница 82: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 82 Appendix Main Menu Structure...
Страница 83: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 83 Review Data Menu Structure...
Страница 84: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 84 Operator Menu Structure...
Страница 85: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 85 Supervisor Menu Structure...