ASTUTE140
®
Meter
ASTUTE140
®
Meter Operation
©Astute Medical, Inc. 2021 PN 300005 Rev G 2021/03/03
48
RFID Cards for Device and Lot Registration
Each new ASTUTE140
®
Electronic Quality Control (EQC) Device, liquid control lot and test lot
must be registered prior to first use. Operators may register ASTUTE140
®
Electronic Quality
Control (EQC) Devices, liquid control lots and test lots only if Supervisors grant registration
permission to Operators in the default settings; otherwise, Supervisors must carry out
registrations. To register an ASTUTE140
®
Electronic Quality Control (EQC) Device, liquid control
lot or test lot, see “ASTUTE140
®
EQC Device Registration” on p. 48, “Liquid Control Lot
Registration” on p. 54, and “Test Lot Registration” on p. 52.
ASTUTE140
®
EQC Device Registration
The EQC procedure verifies the calibration of the ASTUTE140
®
Meter, including the positioning
system, optical system and other internal systems of the meter to confirm that the meter is
functioning properly.
When not in use, the ASTUTE140
®
EQC Device should be stored in its original packaging away
from direct light. If lost, a replacement ASTUTE140
®
EQC Device may be ordered from Astute
Medical, Inc. at www.astutemedical.com or by contacting your Astute Medical, Inc. sales
representative.
The ASTUTE140
®
EQC Device must be registered prior to first use. Operators may register the
ASTUTE140
®
EQC Device only if the Supervisor granted registration permission in the default
settings; otherwise, Supervisors must register the ASTUTE140
®
EQC Device.
To register the ASTUTE140
®
EQC Device, locate the ASTUTE140
®
EQC RFID Card included in
the ASTUTE140
®
Meter Kit and perform the following steps:
1. Press the
key to display the
Main Menu
(if registering the ASTUTE140
®
EQC Device
immediately after successful log in, the
Main Menu
will automatically be displayed).
2. Use the navigation keys to highlight the
Operator Menu
icon.
Figure 52.
Main Menu screen
with Operator Menu
icon highlighted
Содержание ASTUTE140
Страница 1: ...Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 1 ASTUTE140 Meter User Manual...
Страница 82: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 82 Appendix Main Menu Structure...
Страница 83: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 83 Review Data Menu Structure...
Страница 84: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 84 Operator Menu Structure...
Страница 85: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 85 Supervisor Menu Structure...