ASTUTE140
®
Meter
ASTUTE140
®
Meter Operation
©Astute Medical, Inc. 2021 PN 300005 Rev G 2021/03/03
59
8. When prompted, place the prepared Test cartridge into the recess inside the ASTUTE140
®
Meter drawer. Gently close the drawer until it clicks.
9. After closing the drawer, a temporary screen will appear indicating that the ASTUTE140
®
Meter is running the LQC procedure and the amount of time remaining until test completion
will be displayed.
10. When the LQC procedure completes, the results will be displayed along with the word
Passed
if the procedure passed or
Failed
if it did not. If the LQC procedure failed, run the
LQC procedure again following the steps below:
a. Press the right soft key to display the
Options
pop-up menu.
b. Use the
key to highlight
Next Control
and press the right soft key.
c. Repeat steps 4–9 using a freshly prepared Test cartridge. When the LQC procedure
completes, the results will be displayed along with the word
Passed
if the meter
passed the LQC procedure or
Failed
if it did not. If the procedure failed a second
time, contact Technical Support (See “Contacting Technical Support ” on p. 7 for
contact information).
11. Press the
key to the left of the LCD screen to open the ASTUTE140
®
Meter drawer.
Remove the cartridge and discard according to the laboratory’s standard practices for the
type of biological sample used.
12. To manually print the LQC results if Autoprint is disabled:
a. Press the right soft key to display the
Options
pop-up menu.
Figure 64.
Running LQC
Содержание ASTUTE140
Страница 1: ...Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 1 ASTUTE140 Meter User Manual...
Страница 82: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 82 Appendix Main Menu Structure...
Страница 83: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 83 Review Data Menu Structure...
Страница 84: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 84 Operator Menu Structure...
Страница 85: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 85 Supervisor Menu Structure...