Meter Operation
©Astute Medical, Inc. 2021 PN 300005 Rev G 2021/03/03
58
4. A screen showing all liquid control lots previously registered and their expiration dates will
be displayed. Controls displayed as “C” correspond to high liquid controls and those
displayed as “Control –” correspond to low liquid controls.
5. Use the
keys to highlight the correct liquid control lot and level being tested and
press the right soft key. If the liquid control lot is not listed, register the liquid control lot to
be tested by following the steps outlined in “Liquid Control Lot Registration” on p. 54. After
the lot is registered, repeat steps 1–6.
6. After selecting the liquid control lot and level (high or low) to be tested, the ASTUTE140
®
Meter drawer will automatically open.
7. Prepare the Test cartridge with the liquid control (high or low) as instructed in the assay-
specific Liquid Control Kit Package Insert. NOTE: The test lot must be registered prior to
running LQC (See “Test Lot Registration” on p. 52 for instructions).
Figure 63.
Run LQC – Select
Lot screen
Содержание ASTUTE140
Страница 1: ...Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 1 ASTUTE140 Meter User Manual...
Страница 82: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 82 Appendix Main Menu Structure...
Страница 83: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 83 Review Data Menu Structure...
Страница 84: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 84 Operator Menu Structure...
Страница 85: ...ASTUTE140 Meter Appendix Astute Medical Inc 2021 PN 300005 Rev G 2021 03 03 85 Supervisor Menu Structure...