Caution:

 The unit is fitted with a safety mechanism, if the unit encounters an obstruction under 

the front portion of the basin whilst travelling downwards it will automatically stop. Pressing the ‘Up’ 

button will allow the obstruction to be removed. Care should be taken to ensure that 

whilst the basin assembly is moving downwards there are no obstructions left directly 

beneath the basin or rise & fall unit as the safety mechanism is designed to stop the 

unit only when the front of the basin is lifted. 

If both the ‘Up’ and Down’ buttons are pressed at the same time the unit will not move.

Troubleshooting

In the unlikely event that problems occur with the unit, carry out the following easy checks to 

determine whether or not an engineer is required.

Fault:  

The rise and fall unit does not travel up or down

Cause:  

The mains power supply has been disconnected for some time and the unit’s batteries are 

now flat.

Remedy: Examine the green LED fitted to the right hand side of the unit. If the green LED is not 

illuminated neither mains electricity nor battery backup are supplying the unit. Examine the mains 

fuses/circuit breakers and reset as necessary and ensure that mains voltage is fully reconnected 

then press the reset button situated next to the handset connector. If the LED fails to illuminate an 

electrician needs to ensure that mains electrical supply is available at the fused spur (a fuse may have 

blown elsewhere in the circuit).

Fault: 

The rise and fall unit travels upwards but not downwards.

Cause:  

There is an obstruction underneath the basin.

Remedy:  Depress the ‘Up’ button on the handset to gain clearance from the obstruction and then 

remove the obstruction. The rise & fall unit may then resume normal function.

If the problem persists after carrying out these remedies, please

contact Astor-Bannerman on the number on the back page:

Cleaning

The basin and vanity top may be cleaned using most mild cleaning agents. Chemical agents must 

not be used, i.e., drain cleaners, acetone, oven cleaners or acids. The rise and fall unit, mirror and 

handset should be cleaned using a warm soapy solution and rinsed with a damp cloth. Attention 

should always be paid to instruction labels on the outside of cleaning agent containers.

The waste plug

To the rear of the tap block is a small rod with a ‘knob’ fitted to the top. This can be raised and 

lowered to open or close the waste plug as required.

Содержание ABW6

Страница 1: ...Warranty No Height Adjustable Basin ABW6 OWNERS MANUAL INSTALLATION GUIDE ...

Страница 2: ...ould be facing forward towards the user when off To operate them rotate the levers outwards to achieve the flow rate required To stop water flowing rotate the levers until they are facing forward once again Operating the touch tap system The taps are usually set as wave ON Wave OFF To turn the water ON OFF pass your hand above the Sensor on the basin top Touch Tap Sensors in position ...

Страница 3: ... unit Examine the mains fuses circuit breakers and reset as necessary and ensure that mains voltage is fully reconnected then press the reset button situated next to the handset connector If the LED fails to illuminate an electrician needs to ensure that mains electrical supply is available at the fused spur a fuse may have blown elsewhere in the circuit Fault The rise and fall unit travels upward...

Страница 4: ...ble waste assembly including waste plug 1 x three hole mixer tap system 2 x 95mm long 15mm copper pipe 2 x 45mm long 15mm copper pipe 2 x compression elbow 15mm 2 x 15mm straight couplings 2 x 1 2 x 15mm Tap adaptors 2 x lever tap handles 2 x long braided hoses 2 x fibre washers 2 x screen filter seals 1 x 470mm lipped attachment rail 1 x grey 2 button handset 6 x white plastic rivets 1 x mirror 1...

Страница 5: ...equired with flush mounted a flex outlet cover on the wall behind the unit at a height of 1100mm from the floor on the centreline of the unit Fixing The basin assembly should be fixed to a load bearing wall i e strong enough to carry a 100kg load on wall fixing which is flat and free from obstruction from 200mm 2200mm above the floor If the basin needs to be fixed to a stud wall the wall will need specia...

Страница 6: ...ith the one on the wall Ensure that the unit is level then mark the wall through the remaining three attachment points two on the lower horizontal beam and one on the upper horizontal beam Remove the unit from the wall and at the three marked positions drill three 6mm full depth pilot holes Open the two to accept Anchor fixings Open the upper hole to 10mm to accept a plastic wall plug Fit the fixi...

Страница 7: ...ke sure the rod can move freely Using the adjusting screw set the overall length of the basin plug to 77mm Stand the lower cover in place Fit the plug lifting rods in position and set the clamp to lift the plug Touch taps Fit the sensors before mounting the unit on the wall Touch tap sensors should be fitted in the position agreed with the client Drill an 18mm in the position required take care to...

Страница 8: ... brass nut Test the water supply Ensure the isolation valves are CLOSED Place a bucket or suitable container below the plug hole of the basin Open both basin taps fully and then slowly open each of the isolation valves wait 2 minutes then examine the supply system for leaks correct as necessary Close the isolation valves Fit the white plastic lower cover Remove the mushroom heads from the plastic ...

Страница 9: ... supply hoses need to be enclosed within the white self adhesive trunking supplied Using the handset raise the assembly to its highest position Remove the capping from the front of the trunking Remove the protective film from the self adhesive tape fitted to the rear of the trunking Pass the trunking behind the hot and cold supply hoses taking care to ensure that the self adhesive tape is kept awa...

Страница 10: ...Page 10 Version 3 01 04 2016 ...

Страница 11: ...Page 11 Version 3 01 04 2016 ...

Страница 12: ...821110 Technical Helpline 01242 822972 2013 Astor Bannerman Medical Limited Our policy is one of continuous development and we therefore reserve the right to make technical alterations without notice The content of this publication may not be copied either whole or in part without the consent of Astor Bannerman Medical Limited Original Document ...

Отзывы: