Parameterization
VLM60 User Manual
ASTECH GmbH
page 15
PO2Sync
Serves to set output frequency of the second pulse output.
po2s
ync [n]
n = 0, 1
default: 0
0
average-synchronous. The update is synchronous with the duration of the speed averaging
(Parameter
Average
).
1
trigger-synchronous. The update is synchronous with the clock at the switching input (TRIGGER).
PO2Value
Serves to set the output value of the second pulse output.
po2v
alue [c]
c = v, v:x, v:y, a, r
default: v
v
resulting velocity, vector from x- and y-direction
v:x
velocity of x-component
v:y
velocity of y-component
a
angle between the two velocity components, value range: -
180 … +180
· PO2Factor
r
Measurement Rate (Quality of velocity measurement)
Resolution
Serves to set the resolution of the primary sensor.
resol
ution [n]
n = 100, 200, 300, …, 8200
default: 5000
This parameter influences the behavior of the primary sensor. Changes should only be made in consultation
with the manufacturer.
Scantime
Controls the internal temporal resolution of the primary sensor.
scan
time [n]
n = 0 … 65535 (
unit: ms)
default: 0
This parameter influences the behavior of the primary sensor. Changes should only be made in consultation
with the manufacturer.
Silent
Serves to set the behavior of the VLM60 command line (terminal). It can be set whether the entered command
is sent back to the terminal (echo) or not.
silent
[n]
n = 0, 1
default: 0
0
echo on
1
echo off
When using the VLMTool to parameterize the VLM60, this parameter should be set to 0.