Parameter setup and measurement
LDS70A User Manual
Page 50
ASTECH GmbH
TC
–
Time for calibration
Parameterizes the time between 2 customized calibrations.
The calibration is necessary to stabilize the distance accuracy and to avoid a
lower accuracy because of temperature changes in the environmental of the
electronic parts.
Up to a Measuring frequency of approx. 35 kHz the calibration will be done
without any influences of the output frequency.
Between 35 k Hz and 40 kHz it could be possible that the data output will
interrupt for one (1) distance output every x seconds (x= parameter of TC).
TE
–
Terminator
Sets the termination char for the output of measured values in ASCII format.
(see also command SD).
Query:
TC
Set:
TCx
Value range x:
0 (no calibration), 1 … 3660 s (= 1 s to 1 1 h interval)
Standard:
1
Output:
TC 1
Query:
TE
Set:
TEx
Value range x:
0 … 9
Standard:
0
Output:
TE 0