Parameter setup and measurement
LDS70A User Manual
Page 38
ASTECH GmbH
Advantage of GN > 0:
measurements on targets with low reflectivity are
possible
Disadvantage of GN > 0: more interference pulses will lower the accuracy of
the measurement
MF
–
Measurement frequency
MF parameterizes the number of single pulses to be transmitted per second.
This is also the number of single distance measurements per second.
The achievable output frequency depends on the following terms of the setup:
▪
Used interfaces (e.g. serial interface RS232 or RS422, analog output)
▪
length of the interface cable
▪
transfer speed (baud rate
▪
average (parameter SA)
▪
data format of output value (binary, decimal
→
parameter SD)
▪
number of output values (distance, signal strength, temperature
→
parameter SD
Query:
MF
Set:
MFx
Value range:
1 … 40000, resolution 1
Standard:
10000
Example:
Input:
MF 1000
Output:
MF 1000 Hz