Parameter setup and measurement
LDS70A User Manual
Page 40
ASTECH GmbH
Table 7: Decimal output - Maximum measurement frequency (MF
max
)
Output values
Distance
Distance
+ signal quality
Distance
+ temperature
Distance
+ signal quality
+ temperature
Output data
format
SD 0 0
SD 0 1
SD 0 2
SD 0 3
Length of
output data
11
16
17
22
Baud rate
Maximum output frequency = MF
max
with SA 1
9600
80
50
48
40
19200
160
100
90
75
115200
1000
700
660
480
230400
2000
1400
1300
980
460800
4100
2750
2500
1850
921600
7300
4750
4350
3300
2000000
12200
7850
6480
5000
MW
–
Measurement window
The measurement window defines the general distance boundaries for a
successful measurement, starting at distance x and ending at distance y. The
third parameter z sets the output value for the case that a successful distance
measurement results outside the measurement window boundaries.
The measurement window function is useful for:
▪
Blanking out interfering objects before or behind a certain distance
range
▪
Determining a defined measurement range
Query:
MW
Set:
MWx y z
Value range x:
Float32, resolution 0.001 = 1 mm
Value range y
Float32, resolution 0.001 = 1 mm
Value range z
0, 1
Standard:
-71.000 71.000
Output:
MW -71.000 71.000 0