Parameter setup and measurement
LDS70A User Manual
Page 46
ASTECH GmbH
The spread of the measured distance values can be reduced by calculating the
mean average value.
𝜎
𝑆𝐴
=
𝜎
1
√𝑆𝐴
𝜎
𝑆𝐴
Spread after mean value determination from several distance
measurements
𝜎
1
Spread of single measured value (± 60 mm)
𝑆𝐴
Mean value
Example:
Measurement
frequency MF (Hz)
Mean value SA
Output
frequency (Hz)
Spread (mm)
15000
1
15000
± 60
15000
10
1500
± 19
15000
100
150
± 6
15000
1000
15
± 2
Table 9: Mean value outcome
For table above applies:
Measuring frequency 15 kHz and output frequency 15 kHz will be achieved
with baud rate = 921.600 and binary output.
All valid measured distance values will be used for SA. If a distance
value is out of the measurement window MW it will be ignored for
the SA calculation.
SO
–
Set offset
SO carries out a single distance measurement and sets it as - OF (offset). SO
can only be executed and is not really a parameter. SO can be used for zero
adjustment of distances in systems or processes.
Please note the interaction of parameter MW, OF, SO.