Rys. 10. Example of patient’s environment
Non-electromedical device used in electromedical system should comply with
requirements of IEC or ISO safety standards adequate for this particular device.
Devices, in which protection against electric shock depends only on basic insulation,
should not be used in electromedical system,
Responsible divisions should remember, that electromedical system installations
and its modifications during the usage period are the subject to EN 60601- 1
standard requirements.
Responsible division may be: hospital, single doctor or unprofessional person.
In the private usage patient, operator and responsible division may be in one
person.
All non-medical devices, complied with safety standards proper to its original
purposes, and placed in the patients surrounding, require additional appropriations to
reduce touch current, if it exceeds following values:
o
500
µ
A from electromedical system or between such systems in a patients
surrounding in case of cut off of any not installed permanently protective earth
conductor.