Apart from planned (warranty and post-warranty) services of Ergometer, one
should systematically control power supply cable, power supply, ECG cable,
each time before use - it is necessary to check if there are any visible damages.
Any noticed improprieties should be repaired by an authorised ASPEL service
point.
The device is not intended to sterilization.
4.2. Environmental protection
Waste hazardous to the environment is not produced during operation of the
Ergometer CRG 200.
The Ergometer CRG 200 should be utilised when its work life has been over. For
that purpose, it can be sent to the manufacturer or one should contact
a specialist electronic and electromechanic waste company.
4.3. Customer service
If you need help in diagnosing any problems with work of the device please contact
ASPEL S.A. service.
4.4. Dealing with common problems
PROBLEM
SOLUTION
No contact with a PC.
Restart PC.
Restart Ergometer (unplug, wait few seconds, plug again).
Check the cable connection between the Ergometer CRG
200 and a PC.
Replace the old cable with a new one.
If the problem persists, contact an authorised service
representative.
Straight lines on the (ECG
curves) diagram, INOP
message displayed.
Check ECG cable attachment to socket on Egrometer’s
control board.
Check and improve electrodes attachment to patient, using
ECG gel.
If necessary, change place of electrodes contact.
Noises on ECG curves.
Check and improve electrodes attachment to patient, using
ECG gel.
If necessary, turn on digital filtration.