2.4. Functions
Ergometer enables load adjustment within the range of 25 – 1001 W. Rotation
measuring rate is 20 – 150 rpm. Ergometer is controlled by a PC computer through
RS-232 interface.
Load level is controlled by a microprocessor, therefore it does not depend on speed,
which in turn can be adjusted according to patient’s individual needs.
Ergometer is equipped with ECG mode recording 12 standard leads.
Together with Cardio TEST software, Ergometer CRG 200 is used to carry out
exercise tests.
Together with AsTER software, Ergometer CRG 200 it is intended for cardiologic
rehabilitation.
2.5. Main technical and operational parameters
Dimensions (LxWxG):
1030 x 580 x 1240 mm
Weight:
41 kg
Supply:
External power supply CRG 200-M24 (integral part of
Ergometer CRG 200)
Supply voltage:
100 ÷240V AC; 47 ÷ 63 Hz
Max current consumption:
2A (for 100÷120V); 1A (for 220÷240V)
DC socker:
24V; 2A max
Braking:
processor-controlled brake with torque measurement
(load does not depend on speed)
Communication interface:
RS-232
Load level:
25 ÷ 1001 W,
Load accuracy
± 5 W to the power level 50W and ± 10 % above 50 W
Rotational speed:
20 ÷ 150 rpm ± 4 rpm
Max weight of the user:
201 kg
Operation modes:
○
Continuous operation – average load max 100W
○
Continuous operation with intermittent load- average
load max 200 W – working completion factor 66%
(2:1) (maximum working time 30 min; idle time 15
min.)
○
Continuous operation with intermittent load- average
load max 200 W – working completion factor 40%
(2:3) (maximum working time 10 min; idle time 15
min.)
○
Continuous operation with intermittent load- average
load max 200 W – working completion factor 40%
(1:4) (maximum working time 3 min; idle time 12 min.)