- 7 - 

 

Operating modes  

 

This  zone  selects  one  of  the  three  possible  operating  modes.  Pressing  the  button  corresponding  to  the 
mode enables it. The green light comes on continuously to validate the selection. 

Low pressure dynamic mode: enables alternating of one cell out of two (except for the 3 head 

cells). The cycle change is made every 7 minutes.

 

Low  pressure  static  mode:  inflates  all  the  mattress  cells  to  a  constant  pressure  equal  to  the 

lowest  internal  pressure  and  allows  the  patient  to  sink  into  the  support  so  as  to  increase  the  supporting 
area and thus reduce the average interface pressures. 

 Care mode (static): makes the surface of the mattress firm in order to simplify handling of the 

patient during care or when lifting (e.g. transfer from bed to armchair) 

 

 WARNING:   

For safety, the care mode is enabled for 30 minutes; after 25 minutes an audio signal (3 BEEPS) warns that the mode is almost ended 
and the light flashes. At the end of 30 minutes, the compressor reverts automatically to the previous mode.

 

Settings 

This zone is used  to set the  parameters.  Press the  corresponding button to adjust the settings. The 
blue lights come on continuously to validate the selection.

 

 "Pressure" button: this runs a pressure calculation with an AUTOMORPHO compressor at any 

time. 
A steady light means that the calculation is OK. 

 Pressure level indicator: low, medium or high. 

 

 This button adjusts the pressure according to the position of the patient. 

It should be selected for a sitting position greater than 30°. 
  

HINT: 

In some cases, if the slope of the bed chest elevator is greater than 30°,  you may have to increase the pressure inside the cells for the 

comfort and safety of the user 

 Press the 

 key and a blue light comes on.  

 

We recommend that you limit the time in the semi-seated position at 45° to a maximum of 2 hours to reduce the risk of eschar. 

 

Alarms 

This zone is devoted to alarms:

 

  Audio-visual  alarm  indicating  abnormal  pressure:  the  indicator  flashes  and  a  bell  sounds  if  the 

pressure is too low or too high and in this way warns of leaks or pump failures. 

  Audio-visual  alarm  indicating  an  electrical  fault:  the  indicator  flashes  and  a  bell  sounds  if  the 

power supply is cut-off or the power lead is disconnected or altered.

 

 

 
 
 
 

 

 

 

 

Содержание WINNCARE automorpho AUTOMORPHO

Страница 1: ... PA0010085 MAJ 08 01 2013 UK User manual ...

Страница 2: ......

Страница 3: ......

Страница 4: ...CTOGRAMS 3 C DESCRIPTION 3 D HOW IT WORKS 3 E INDICATIONS 4 F PRECAUTIONS 4 G SPECIFICATION 5 H CONTROL PANEL 6 I INSTALLATION USE 8 J ALARMS 9 K MAINTENANCE DISINFECTION 10 L MAINTENANCE SERVICING 10 M STORAGE 12 N DISPOSAL OF THE MEDICAL DEVICE 12 O WARRANTY 12 ...

Страница 5: ...foam base The cells are inflated alternately one out of two except for the 3 head cells which remain static The whole is protected by an integral removable waterproof and breathing cushion incorporating a pillow D HOW IT WORKS The alternating pressures prevent prolonged vascular compression which may lead to histohypoxia The low pressure static mode enables you to handle patients who need to be im...

Страница 6: ... European Directive 93 42 CEE relating to the essential criteria applicable to medical devices only compatibility between systems assembled by the manufacturer ASKLESANTE guarantees a safe combination when using the AXTAIR motorised air mattress The specification and performance of the motorised air support will only be maintained by using the AXTAIR pump ref VAXT4 POMPE AUTO associated with the A...

Страница 7: ...d inflation pump Compatible ref VKIT AXT SPECIFICATION OF THE AUTOMORPHO COMPRESSOR Alternating mode 1 cell out of 2 Alternates every 7 minutes Low pressure static mode Automatic pressure control dependent upon the patient s morphology Manual optimisation in semi seated position 50 Static mode for care closed plan Flow 8 litres min Mattress inflation time Less than 20 mins for an Automorpho Approx...

Страница 8: ...initial inflation Servicing indicator This orange indicator lights up continuously if servicing of the compressor is advisable The indicator flashing together with an alarm sound indicates a problem with the distributor In this case please contact your reseller See L MAINTENANCE SERVICING Mains indicator This orange indicator when on permanently indicates that the compressor is connected to the ma...

Страница 9: ...ettings This zone is used to set the parameters Press the corresponding button to adjust the settings The blue lights come on continuously to validate the selection Pressure button this runs a pressure calculation with an AUTOMORPHO compressor at any time A steady light means that the calculation is OK Pressure level indicator low medium or high This button adjusts the pressure according to the po...

Страница 10: ...ndicators come on alternately For a precise calculation the patient should not move excessively The calculation lasts a maximum of 6 minutes When the calculation is finished the pressure button indicator goes out and the pressure indicators are lit continuously to indicate that the pressure is now matched to the patient s morphology The compressor then starts DYNAMIC mode automatically 11 Selectin...

Страница 11: ...onnect the air lines blank them off with the transport plug and move the patient regularly When the power supply returns you must restart the compressor by pressing the Start Stop button while repeating the INSTALLATION USE instructions I starting from step 8 after you have reconnected the mattress feed pipe to the compressor COMMENTS On starting the compressor there is an initialisation step If t...

Страница 12: ...nence time The removable foam base can be replaced in case of alteration Ref VAXT3 MS support x4 Base 5cm CE marking compulsory under directive 93 42 CEE for products claiming that they are suitable for the surfaces of Medical Devices Cover Observe the following washing conditions Wash in water maximum temperature 90 C reduced mechanical action rinsing in decreasing temperature spin drying Bleachi...

Страница 13: ...ré Traitement de surface Décontamination Répartition de surface par µfibres Solution base alcoolique EPI masque Hygiène gants tablier Propreté physique Inacceptable Entretien de surface Méthode 1 Opération Produit Détergent Désinfectant EPI gants tablier Lavage housse en machine à 90 C Détergent Housse de remplacement identique à la protection d origine Contrôle Qualité Conforme Matelas et Compres...

Страница 14: ...ctronic equipment waste at the end of its working life You can contact the store reseller or distributor from whom the product was bought to find out the collection points for worn out devices closest to your home Before any disposal the appliance must be cleaned in accordance with the instructions in K SERVICING DISINFECTION to avoid any risk of contamination We are grateful for your co operation...

Страница 15: ......

Страница 16: ...ASKLÉSANTÉ 200 rue Charles Tellier ZI du Grézan 30 034 Nîmes Cedex 1 France Tél 33 0 4 66 02 15 15 Fax 33 0 4 66 02 15 00 Email contact askle com www askle com ...

Отзывы: