INDICATIONS 

 

In  the  opinion  of  the  Committee  for  Evaluation  of  Goods  and  Services  published  in  the 

Journal  Officiel

  on  21  July  2005  and  in  the 

opinion of clinical experts: 
 
Prevention and aid to treatment of eschars from stages 1 to 4 (according to medical opinion) in patients confined to bed for  more 
than 15 hours, presenting with a risk of eschars assessed against a validated scale and clinical judgment. 
Patients whose weight cannot be determined. 

 
 
 

F

 

PRECAUTIONS 

 

Non-stabilised and/or muscular trauma in contact with the support; 
Days immediately following surgical intervention on eschar (skin graft or pedicled graft) [

 prefer static mode]; 

Patient monitored at home without the possibility of intervention by medical auxiliaries; 
The device must not be used in a hyperbaric chamber. 

 

 WARNING: 

According  to  appendix  1  to  European  Directive  93/42/CEE  relating  to  the  essential  criteria  applicable  to  medical  devices,  only 
compatibility  between  systems  assembled  by  the  manufacturer, 

ASKLE

SANTE,  guarantees  a  safe  combination  when  using  the 

AXTAIR motorised air mattress. 

 

The  specification  and  performance  of  the  motorised  air  support  will  only  be  maintained  by  using  the  AXTAIR  pump  [ref. 
VAXT4/POMPE/AUTO],  associated  with  the  AXTAIR  mattress  [ref.  VAXT4/MA/AUTO  or  VAXT4/MA/XL]  and,  optionally,  the 
inflation/deflation kit [ref. VKIT/AXT].

 

The national authority responsible for health and safety and health products can at any time institute audits of the conditions under 
which said products are marketed and take the measures necessary if there is danger or a breach of the rules. In the event of failure 
to comply with the abovementioned conditions of use, the user's liability may be committed if there is an accident.  

  PLEASE  NOTE

:

  The  compressor  and  the  mattress  are  still  compatible  with  the  products  with  equivalent 

references from previous ranges

 

 

 

Содержание WINNCARE automorpho AUTOMORPHO

Страница 1: ... PA0010085 MAJ 08 01 2013 UK User manual ...

Страница 2: ......

Страница 3: ......

Страница 4: ...CTOGRAMS 3 C DESCRIPTION 3 D HOW IT WORKS 3 E INDICATIONS 4 F PRECAUTIONS 4 G SPECIFICATION 5 H CONTROL PANEL 6 I INSTALLATION USE 8 J ALARMS 9 K MAINTENANCE DISINFECTION 10 L MAINTENANCE SERVICING 10 M STORAGE 12 N DISPOSAL OF THE MEDICAL DEVICE 12 O WARRANTY 12 ...

Страница 5: ...foam base The cells are inflated alternately one out of two except for the 3 head cells which remain static The whole is protected by an integral removable waterproof and breathing cushion incorporating a pillow D HOW IT WORKS The alternating pressures prevent prolonged vascular compression which may lead to histohypoxia The low pressure static mode enables you to handle patients who need to be im...

Страница 6: ... European Directive 93 42 CEE relating to the essential criteria applicable to medical devices only compatibility between systems assembled by the manufacturer ASKLESANTE guarantees a safe combination when using the AXTAIR motorised air mattress The specification and performance of the motorised air support will only be maintained by using the AXTAIR pump ref VAXT4 POMPE AUTO associated with the A...

Страница 7: ...d inflation pump Compatible ref VKIT AXT SPECIFICATION OF THE AUTOMORPHO COMPRESSOR Alternating mode 1 cell out of 2 Alternates every 7 minutes Low pressure static mode Automatic pressure control dependent upon the patient s morphology Manual optimisation in semi seated position 50 Static mode for care closed plan Flow 8 litres min Mattress inflation time Less than 20 mins for an Automorpho Approx...

Страница 8: ...initial inflation Servicing indicator This orange indicator lights up continuously if servicing of the compressor is advisable The indicator flashing together with an alarm sound indicates a problem with the distributor In this case please contact your reseller See L MAINTENANCE SERVICING Mains indicator This orange indicator when on permanently indicates that the compressor is connected to the ma...

Страница 9: ...ettings This zone is used to set the parameters Press the corresponding button to adjust the settings The blue lights come on continuously to validate the selection Pressure button this runs a pressure calculation with an AUTOMORPHO compressor at any time A steady light means that the calculation is OK Pressure level indicator low medium or high This button adjusts the pressure according to the po...

Страница 10: ...ndicators come on alternately For a precise calculation the patient should not move excessively The calculation lasts a maximum of 6 minutes When the calculation is finished the pressure button indicator goes out and the pressure indicators are lit continuously to indicate that the pressure is now matched to the patient s morphology The compressor then starts DYNAMIC mode automatically 11 Selectin...

Страница 11: ...onnect the air lines blank them off with the transport plug and move the patient regularly When the power supply returns you must restart the compressor by pressing the Start Stop button while repeating the INSTALLATION USE instructions I starting from step 8 after you have reconnected the mattress feed pipe to the compressor COMMENTS On starting the compressor there is an initialisation step If t...

Страница 12: ...nence time The removable foam base can be replaced in case of alteration Ref VAXT3 MS support x4 Base 5cm CE marking compulsory under directive 93 42 CEE for products claiming that they are suitable for the surfaces of Medical Devices Cover Observe the following washing conditions Wash in water maximum temperature 90 C reduced mechanical action rinsing in decreasing temperature spin drying Bleachi...

Страница 13: ...ré Traitement de surface Décontamination Répartition de surface par µfibres Solution base alcoolique EPI masque Hygiène gants tablier Propreté physique Inacceptable Entretien de surface Méthode 1 Opération Produit Détergent Désinfectant EPI gants tablier Lavage housse en machine à 90 C Détergent Housse de remplacement identique à la protection d origine Contrôle Qualité Conforme Matelas et Compres...

Страница 14: ...ctronic equipment waste at the end of its working life You can contact the store reseller or distributor from whom the product was bought to find out the collection points for worn out devices closest to your home Before any disposal the appliance must be cleaned in accordance with the instructions in K SERVICING DISINFECTION to avoid any risk of contamination We are grateful for your co operation...

Страница 15: ......

Страница 16: ...ASKLÉSANTÉ 200 rue Charles Tellier ZI du Grézan 30 034 Nîmes Cedex 1 France Tél 33 0 4 66 02 15 15 Fax 33 0 4 66 02 15 00 Email contact askle com www askle com ...

Отзывы: