MAINTENANCE/DISINFECTION 

 

 RISK OF ELECTRIC SHOCK:  

 

Before any maintenance, disconnect the equipment from its power supply in order to avoid any risk of physical injury or damage to 
equipment. 

 

Never expose the equipment to sprayed water.

 

 

 RISK OF DAMAGE TO THE EQUIPMENT:  

 

You must not use corrosive cleaning products such as industrial degreasing agents, solvents of the acetone type and ether.  

 

You must not use abrasive materials such as steel wool or "

ScotchBrite".

 

 

Compressor: 

 

 
Use  a  cloth  lightly  soaked  with  a  detergent  or  surface  detergent/disinfectant  solution  marked  CE*  at  the  concentrations  for  use 
recommended by the manufacturer. Observe the remanence time. 

Please note: In order to prevent any damage to the compressor through liquid getting inside the casing, please do 
not spray the product 

 

 
 

Mattress: 

 
Disinfection  by  hand:  use  a  cloth  impregnated  with  a  solution  of  detergent  or  surface  detergent/disinfectant  marked  CE*  at  the 
concentrations for use recommended by the manufacturer. Observe the remanence time.

 

The removable foam base can be replaced in case of alteration [Ref. VAXT3/MS "support x4 Base>5cm"] 

 
* CE marking compulsory under directive 93/42/CEE for products claiming that they are suitable for the surfaces of Medical Devices. 

 

Cover: 

Observe the following washing conditions: 

 

 

Wash  in  water,  maximum  temperature  90°C,  reduced  mechanical  action,  rinsing  in  decreasing 
temperature, spin drying. 

 

Bleaching possible, chlorination at 1000 ppm permitted. 

 

Do not iron. 

 

Do not dry clean or use solvent-based stain removers. 

 

Tumble drying permitted, moderate temperatures 

  
The  MANUFACTURER  will  provide  on  request  wiring  diagrams,  lists  of  components,  descriptions,  calibration  instructions  or  any 
other relevant information to SERVICING and MAINTENANCE STAFF in order to carry out permitted repairs in accordance with 
the contract between the applicant and ASKLE, the manufacturer. 

 

 
 

L

 

MAINTENANCE/SERVICING 

 
Periodic servicing of AUTOMORPHO AXTAIR compressors is advised in order to maintain an ideal operating specification.  

The periodic servicing interval recommended is indicated by the servicing light on the control panel coming on. 

 

 

This servicing can only be carried out by authorised persons; 

contact your reseller. 

 

Содержание WINNCARE automorpho AUTOMORPHO

Страница 1: ... PA0010085 MAJ 08 01 2013 UK User manual ...

Страница 2: ......

Страница 3: ......

Страница 4: ...CTOGRAMS 3 C DESCRIPTION 3 D HOW IT WORKS 3 E INDICATIONS 4 F PRECAUTIONS 4 G SPECIFICATION 5 H CONTROL PANEL 6 I INSTALLATION USE 8 J ALARMS 9 K MAINTENANCE DISINFECTION 10 L MAINTENANCE SERVICING 10 M STORAGE 12 N DISPOSAL OF THE MEDICAL DEVICE 12 O WARRANTY 12 ...

Страница 5: ...foam base The cells are inflated alternately one out of two except for the 3 head cells which remain static The whole is protected by an integral removable waterproof and breathing cushion incorporating a pillow D HOW IT WORKS The alternating pressures prevent prolonged vascular compression which may lead to histohypoxia The low pressure static mode enables you to handle patients who need to be im...

Страница 6: ... European Directive 93 42 CEE relating to the essential criteria applicable to medical devices only compatibility between systems assembled by the manufacturer ASKLESANTE guarantees a safe combination when using the AXTAIR motorised air mattress The specification and performance of the motorised air support will only be maintained by using the AXTAIR pump ref VAXT4 POMPE AUTO associated with the A...

Страница 7: ...d inflation pump Compatible ref VKIT AXT SPECIFICATION OF THE AUTOMORPHO COMPRESSOR Alternating mode 1 cell out of 2 Alternates every 7 minutes Low pressure static mode Automatic pressure control dependent upon the patient s morphology Manual optimisation in semi seated position 50 Static mode for care closed plan Flow 8 litres min Mattress inflation time Less than 20 mins for an Automorpho Approx...

Страница 8: ...initial inflation Servicing indicator This orange indicator lights up continuously if servicing of the compressor is advisable The indicator flashing together with an alarm sound indicates a problem with the distributor In this case please contact your reseller See L MAINTENANCE SERVICING Mains indicator This orange indicator when on permanently indicates that the compressor is connected to the ma...

Страница 9: ...ettings This zone is used to set the parameters Press the corresponding button to adjust the settings The blue lights come on continuously to validate the selection Pressure button this runs a pressure calculation with an AUTOMORPHO compressor at any time A steady light means that the calculation is OK Pressure level indicator low medium or high This button adjusts the pressure according to the po...

Страница 10: ...ndicators come on alternately For a precise calculation the patient should not move excessively The calculation lasts a maximum of 6 minutes When the calculation is finished the pressure button indicator goes out and the pressure indicators are lit continuously to indicate that the pressure is now matched to the patient s morphology The compressor then starts DYNAMIC mode automatically 11 Selectin...

Страница 11: ...onnect the air lines blank them off with the transport plug and move the patient regularly When the power supply returns you must restart the compressor by pressing the Start Stop button while repeating the INSTALLATION USE instructions I starting from step 8 after you have reconnected the mattress feed pipe to the compressor COMMENTS On starting the compressor there is an initialisation step If t...

Страница 12: ...nence time The removable foam base can be replaced in case of alteration Ref VAXT3 MS support x4 Base 5cm CE marking compulsory under directive 93 42 CEE for products claiming that they are suitable for the surfaces of Medical Devices Cover Observe the following washing conditions Wash in water maximum temperature 90 C reduced mechanical action rinsing in decreasing temperature spin drying Bleachi...

Страница 13: ...ré Traitement de surface Décontamination Répartition de surface par µfibres Solution base alcoolique EPI masque Hygiène gants tablier Propreté physique Inacceptable Entretien de surface Méthode 1 Opération Produit Détergent Désinfectant EPI gants tablier Lavage housse en machine à 90 C Détergent Housse de remplacement identique à la protection d origine Contrôle Qualité Conforme Matelas et Compres...

Страница 14: ...ctronic equipment waste at the end of its working life You can contact the store reseller or distributor from whom the product was bought to find out the collection points for worn out devices closest to your home Before any disposal the appliance must be cleaned in accordance with the instructions in K SERVICING DISINFECTION to avoid any risk of contamination We are grateful for your co operation...

Страница 15: ......

Страница 16: ...ASKLÉSANTÉ 200 rue Charles Tellier ZI du Grézan 30 034 Nîmes Cedex 1 France Tél 33 0 4 66 02 15 15 Fax 33 0 4 66 02 15 00 Email contact askle com www askle com ...

Отзывы: